Digital Tesla Meter Gauss Meter Field Strength Meter for Magnetic Particle Inspection HGS-10A
 
This instrument is the magnetic field measuring instrument specially designed for field measurement. Its design concept of lightness, practicality and portability has been determined at the beginning of instrument design and its unique design of micro power consumption allows its battery life to be greatly improved.

 

I Main characteristics

  • With small volume, light weight and convenient carry, this instrument can meet the measurement requirements of various sites;
  • Adopt micro power consumption design. With long service life, the battery can be continuously used for more than 50 hours;
  • Can select transverse probe or axial probe;
  • Have judgment function of magnetic field polarity.

 
II Main technical index

  • Supply voltage: One 9V laminated cell
  • Display meter: 4 1/2 digital display
  • Measurement scope:0~200mT,0~2000mT
  • Accuracy: 2%
  • Resolution: 0.01mT
  • Temperature error: -0.06%/°C
  • Operating environment: Temperature: -10°C~40°C, humidity: 35~75%
  • Backlight YES
  • Peak value hold YES

III Instrument operating instruction
 

Step 1: Connection: connect the Hall probe with the instrument⑥;
Step 2: Power on: Press the “POWER” switch ②on the panel of handhold Gauss meter and there shall be display ①on the panel meter, it represents the instrument power has been energized and measurement can be done;
Step 3: After the power is energized, firstly select a corresponding range ③according to the magnetic field size of the measured object.
Step 4: Zero point correction: Place the Hall probe in the area whose magnetic field is zero. When the meter reading is not zero, adjust “Zero” ⑤potentiometer till the reading is zero.
Step 5: Magnetic field measurement
A. Place the Hall probe in the surface magnetic field of the measured sample, move the Hall probe and the magnetic field intensity can be directly read from the instrument, whose unit is mT and the magnetic field polarity also can be judged (When using transverse probe and the display value is “-”, the number side of the Hall probe faces N pole and when using axial probe and display value is “-”, the end side of the Hall probe faces N pole);
B. The range button, the range is 2000mT when it is pressed and the range is 200.0mT when it bounces back;
C. Contact the magnetic steel surface with Hall chip side in the practical measurement, the measurement result is closer to the actual value. When measuring with base plate side, the measurement result is little low, 1mT = 10 Gs.
 

 

Standard delivery

Main unit

Transverse fiber probe

Instruction manual

Calibration certificate

Carrying case

Power adaptor

 

Optional accessories

Lognitudinal probe

Transverse metal probe

 

Hot Products

Ferrite meter Ferrite content tester Ferrite tester HUATEC HFE-100 SP10A 1. Application In the ...
Ferrite content measurements 1. Preface In the chemical industry the stress-bearing members of ...
Thru Coating high precision digital Ultrasonic Thickness Gauge TG-3250 TG-3250 is a high precision ...
Ultrasonic Through Coating Thickness Gauge TG4100,5MHz Echo To Echo TECHNICAL DATA DISPLAY 128X64 ...
Industry X ray Film Processor Developer ,Film Developer Machine TQ-14TZ TQ-14 series fully automatic ...
High precision 75 degree Paper Gloss Meter HGM-BZ75 for paper testing Gloss Meter Model HGM-BZ75 (75...