Beijing Zetron Technology Co., Ltd |
Verified Suppliers
|
|
UV-200AT Ozone Concentration Analyzer
UV-200AT benchtop portable ozone concentration analyzer (negative
pressure sampling) is mainly used for ozone concentration
measurement in the ozone generator pipeline, and can continuously
detect the ozone concentration in the ozone generator outlet
pipeline online.
This ozone gas analyzer is designed and manufactured according to
Lambert-Bier's law, the National Environmental Standard of the
People's Republic of China "Determination of Ozone in Ambient Air -
Ultraviolet Photometric Method", the Ministry of Construction of
the People's Republic of China "Ozone Generators for Water
Treatment", and the test and calibration methods. By measuring the
ozone absorption of ultraviolet light before and after the light
signal intensity changes, thus calculating the current ozone
concentration value.
The equipment uses advanced dual-light UV light source system with
lamp intelligent management system, which can make the UV lamp
quickly reach the measurement state after power on, using foreign
separated light cell technology, the equipment has no leakage, high
pressure resistance, high flow sampling gas impact resistance, easy
to clean, easy maintenance, simple operation, low cost of use, etc.
The equipment automatically zero blowing once every certain time to
ensure the stability of the zero point, to prevent the zero point
drift affect the accuracy of the detection concentration, the gas
flow, pressure and flow rate remain unchanged during the
calibration zero, to prevent the impact of airflow on the
calibration zero, to ensure that the equipment can operate
continuously 24 hours a day.
1. Measurement principle: accurate measurement by photometric
absorption principle based on Lambert-Beer's law.
2. Test method: dual-optical UV absorption method, high measurement
accuracy.
3. Light source system: imported long-life UV lamps (wavelength
253.7nm).
4. Usage: can be installed and used outdoors, the instrument is
waterproof, dustproof, and can run continuously for 24 hours,
equipped with waterproof power supply and signal connector.
5. Light cell system: foreign separated light cell technology, no
leakage, high pressure resistance, resistance to high flow sampling
gas impact.
6. Intelligent compensation: built-in light source automatic
compensation function.
7. Automatic zero calibration: every certain time the device
automatically zero calibration 1 time, you can choose the manual
automatic mode to ensure the accuracy of the detection
concentration.
8. Data display: high-definition digital backlight display, clear
and countable data.
9. Output function: 4-20mA current signal, RS485 communication.
UV-200AT is mainly used for continuous detection of ozone gas in various industrial environments, and can also be used to detect ozone gas in operating pipelines, vessels and other environments. It is widely used in pharmaceutical, chemical, municipal and wastewater treatment industries to measure the ozone generator outlet concentration and ozone generator output.
Range | 0-200g/m3NTP (optional range: 0-100g/m3, 0-200g/m3, 0-250g/m3) |
Display resolution | 0.01 g/m3 |
Gas flow | 0.8L±0.2L/min |
Input pressure | <0.1 MPa |
Zero point drift | <±0.3%.FS (MAX: ±0.6g/m3) |
Sampling port pipe diameter | Φ6 (6mm*4mm) |
Error rate | ±0.5% |
Communication mode | RS-485 |
Output mode | 4-20mA, linear |
Size | 260mm (L) × 120mm (W) × 230mm (D) |
Tag: Exhaust ozone detector, Ozone detector, Ozone concentration online monitor, Ozone generator manufacturer, Tabletop ozone analyzer,Wall-mounted Ozone Gas Analyzer, OEM/ODM services