Aluminum Alloy R Axis Optical Rotation Stage 82mm Diameter

Country/Region china
City & Province beijing beijing
InquireNow

Product Details

Optical Sliding Table , R Axis Manual Rotation Stage, Rotating Platform PT-SD206

 

Precision manual rotary table: PT-SD206

 

Product features:

 

■ the table can be rotated 360 ° for rough adjustment.

 

■ fine adjustment of ± 10 °, micrometer head drive, spring return, high resolution and no empty return.

 

■ precision shafting design, adjustable shafting clearance and stability.

 

■ 360 ° laser marking around the circumference for easy reading.

 

■ there are connecting holes with standard hole spacing at the upper and lower parts, which is convenient to connect with other displacement tables to form a multi-dimensional mobile table.

 

Product parameters:

 

Item Number

PT-SD206

Rotation Range

coarse 360° fine ±10°

Diameter of Platform:

Ø82mm

Actuator Type

Micrometer Head in Side

Graduations

Minimum Readout

2'
Minimum Adjust Distance15"

Surface Roundness

60μm
Transversal Deviation30μm

Parallelism

60μm

Material – Finish

Aluminum Alloy - Black-Anodized

Load Capacity

5Kgs

Weight

0.5kgs

 

Product Dimension Map:

Rotary table center correction

 

The purpose is to make the vertical axis of the rotating table coincide with the microscope axis. Move the sheet so that a small mineral particle (hereinafter referred to as particle) is placed on the crosshair intersection, each axis is at the zero position, and the stage is locked at 0 °.

 

Rotate the n-axis 360 °, if the particle makes a circular motion around the center of the crosshair, it indicates that the center is not correct and needs to be corrected. The principle of the method is the same as that of the correction of the objective lens center of the microscope. Rotate the n-axis 180 °, and the particle is the farthest from the center. Gently rotate the two correction screws of the additional chassis (if there is no such chassis, gently push the base that has not been locked with both hands), move the particle to half the center of the crosshair, and move the sheet by hand to place the original particle or another particle in the center of the field of view.

 

Turn the n-axis inspection again. If the particle still moves away from the center of the crosshair, adjust the correction screw of the rotating table (or move the base) again until the particle in the field of view center does not deviate from the center of the crosshair during the n-axis inspection. For the rotary table without additional chassis, be sure to lock the base fixing screw at this time.

 

Hot Products

High Performance Vacuum Motorized Linear Stage Can Be Customized The product does not include the ...
Vacuum Electric Control Linear Stage Precision Linear Slider Guide Translation Stage Product ...
Electric Two Axis Stretching Stage High Precision Motorized Linear Stage Stretching electric linear ...
XYZ Three-Axis Displacement Satge Micro Motorized Linear Stage Product description: This motorized ...
Fine Grinding Lead Screw Motorized Linear Stage Thin And Light Precision Motorized Linear Stage: ...
Light Load Low-Speed High-Resolution Electric Linear Platform Precision Electric Linear Platform: ...