Home Companies Zhejiang Shengdi Technology Inc.

HS-9521B Single Phase Isolated Current Transformer(For Single Meter Test Bench)

Zhejiang Shengdi Technology Inc.
Active Member

Contact Us

[China] country

Address: #8,Zhenghua Road,Wuyuan Industry Zone,Haiyan County

Contact name:Yeli Lu

Inquir Now

Zhejiang Shengdi Technology Inc.

HS-9521B Single Phase Isolated Current Transformer(For Single Meter Test Bench)

Country/Region china
InquireNow

Product Details

HS-9521B single phase ICT

Is used on multi position test benches for testing single phase meters with closed links between the current and voltage measuring circuits (CP links), to increase the work efficiency. Meter of this type are produced and used with increasing frequency.

ICT is used to avoid unnecessary connection between current and voltage circuit in single phase test benches when testing closed links watt-hour meters or when testing opened links watt-hour meters, current and voltage circuit are not likely to be separated for increasing work frequency.

compensation performance .Its phase error, amplitude error are tiny and won’t bring significant additional error. 

Advantage:

• Wide-range: 10mA—120A;

• Output power (Max.):40VA;

• High accuracy class 0.05 by electronic error compensation;

• Protection against  Open circuit and Over load;

• Automatic unused position by-pass;

• Remote operation and control;

 

Application:

• On multi position test systems for meters with closed current-voltage links;

• Modify the existing testing devices

 

General Characteristics:

Voltage:  ±24VDC;

Max. Power Consumption:max.15VA;

Outer: Aluminum case;

Product Size: 180*140*200mm(W*D*H);

Hole Size: φ36mm;

Weight: 7.5kg;

Working Temperature:     -10°C ---- +50°C;

Temperature Coefficient: ≦0.005%/°C( -10°C ---- +50°C);

 

CT Characteristics:

Working Range:10mA—120A;

Frequency:    45--65 Hz;

Current Ratio: 1:1(primary current = secondary current);

Output Power

 

Current

120A

100A

80A

60A

10A

1A

100mA

Max. Output Power

60VA

40VA

32VA

30VA

4VA

40mVA

0.4mVA

Max. Power Consumption

5VA

3.5VA

2.2VA

1.3VA

(ignore)

Input Load

0.35mΩ

 

Output Load

1A … 120A

100mA … 1A

Current

120A

100A

80A

60A

10A

1A

100mA

Max. Load

4.2 mΩ

4 mΩ

5mΩ

8.3 mΩ

40 mΩ

40 mΩ

40 mΩ

Output Load Voltage

0.4V

40 mΩ*I

 

Error

 

Testing Current

100mA—120A

25mA—100mA

10mA—25mA

Ratio Error

≦±0.02%(Normal);

≦±0.05%(Max.);

≦±0.10%(Normal);

≦±0.20%(Max.);

≦±0.50%(Normal);

Angle Error

≦±0.8min

≦±1.5min

≦±3min

Testing Range

cosΦ=0.5c---1---0.5L

 

Green LED(OK):

the phase ICT operating normally;

 

Red LED(OVL):

the phase ICT is under protection;

SHORT key:

Short ICT to protection manually ;

RESET key:

Reset ICT to protection manually ;

 

Contact: Yeli

Email: hysdsales@gmail.com

Skype: lily19850618

 

 

 

Hot Products

HS-3103 1 Position Portable Single Phase Meter Test Equipment Suit for Lab Output voltage and ...
HS3103 Portable 1 Phase Energy Meter Test Equipment Technical specification 1. Electrical output 1) ...
HS-3103 Portable Single Phase Energy Meter Test Bench Output voltage and current distortion not ...
HS3303 Portable 3 Phase Energy Meter Test Bench Technical specification 1. Electrical output 1) ...
HS3303 Three Phase Energy Meter Test Bench 1.with Three-phase Standard Energy Meter 2.Can hung with ...
Output voltage and current distortion not larger than 0.3%(if it is a linear load) The variation of ...