UV Sterilization 100 Level Protection For Chemical Laboratories BSC Biosafety Cabinet

BSC Biosafety Cabinet Product Description:

The safety cabinet is the main isolation equipment used in the laboratory, which can prevent the diffusion of harmful suspended particles and provide protection for the experimenter, the experiment object and the environment.

◆ Air curtain isolation design prevents internal and external cross-contamination. (Negative pressure type)

◆ The glass door moves up and down, equipped with a height limit alarm prompt, so that the operator can use it correctly.

◆ Special filters are installed at the exhaust to control emission pollution.

◆ Equipped with a dedicated waterproof socket to provide great convenience for the operator.

◆ Using LED LCD panel control.

◆ The experimental area is made of high-quality 304 stainless steel, and the working area is equipped with a 2KW spare socket to provide convenience for the operator. Special note: This product is not a medical device.

 

ModelBHC-1000A2
Overall size1200×750×1950mm
Purification area size1000×650×580mm
 

 

The working principle of UV Sterilization 100 Level Protection For Chemical Laboratories BSC Biosafety Cabinet:

The working principle of the biological safety cabinet is mainly to suck the air inside the cabinet to maintain a negative pressure in the cabinet and protect the staff through the vertical airflow; the outside air is filtered by a high-efficiency particulate air filter (HEPA) After filtering, enter the safety cabinet to avoid contamination of the processed samples; the air in the cabinet also needs to be filtered by a HEPA filter before being discharged into the atmosphere to protect the environment.

Installation of BSC Biosafety Cabinet:

1.  The biological safety cabinet shall not be placed on its side, impacted, or collided during transportation, and shall not be directly attacked by rain, snow, or sun exposure.

2.  The working environment of the biological safety cabinet is 10~30℃ and the relative humidity is <75%.

3.  The equipment should be installed on a horizontal surface that cannot be moved.

4.  The equipment must be installed close to a fixed power socket. When the external exhaust system is not connected, the top of the device must be at least 200mm away from the obstacle at the top of the room, and the distance between the back and the wall must be at least 300mm to facilitate the smooth flow of external exhaust. Maintenance of safety cabinets.

5.  In order to prevent airflow interference, it is required not to install the equipment in the passage of people, and not to make the operating port of the sliding front window of the biological safety cabinet face the door and window of the laboratory or be too close to the door and window. Places that may disturb the airflow.

6.  When used in high altitude areas, the wind speed must be recalibrated after installation.

 

Hot Products

TG16MW Normal Temperature Desktop No Noise High Speed Centrifuge TG16MW High Speed Centrifuge: It is ...
TG16MW High Speed Centrifuge: It is widely used in experiments with small sample volume, many steps, ...
TG16MW Safe Ultra-Quiet Research Use High Speed Refrigerated Centrifuge The high speed refrigerated ...
20800r/Min Lab High Speed Centrifuge Machine Continuous Run High Speed Centrifuge Machine Features: ...
12000r/Min High Speed Refrigerated Centrifuge All Steel Body Features of High Speed Refrigerated ...
TG16MW Normal Temperature Desktop No Noise High Speed Centrifuge TG16MW High Speed Centrifuge: It is ...