Vertical Autoclave Sterilizer Features:
◆ Auto discharge the cool air, and steam discharging automatically
after sterilization
◆ Automatically shut off with beep reminding after sterilization
◆ The accident protection system
Model | Material | Volume | Power Supply | Output power | Dimension of sterilizing chamber | Net weight | Package Size |
XFH-50CA | All stainless steel | 50L | 220V | 3.5kW | 340 x 525mm | 58Kg | 570 x 570 x 1300mm |
Vertical autoclave sterilizer use and maintenance instructions:
1. Vertical autoclave sterilizer, the operation staff must be
in the correct use of standardized operation, to avoid an accident.
2. should ensure that container of water, high water level
can make the wet dressing, and waste of power supply, water level
is too low will not work properly.
3. began to heat, must be steam feed valve handle in
position, make the cold air inside the container escape, otherwise
you will notTo the disinfection effect of the good.
4. after each terminated after disinfection sterilization
must shut off the power switch.
5. The solution of disinfection sterilization, the solution
should be pumped into the high temperature resistant glass bottles,
injection quantity for the bottle of 34 volume advisable,
application gauze on the bottle, avoid by all means use a stuffy
cork, and the solution bottle placed in a container can hold
solution quantity, in case of a one thousand solution bottle burst
in the process of disinfection sterilization, and other items.
6. after the termination of disinfection vertical autoclave
sterilization, disinfection fungus is a solid objects, but will
bleed valve handle in the vent steam, when the pressure gauge
pointer back to "0" position, can open the container cover, if
disinfection sterilization items solution or medium etc., do not
immediately release the steam, solutions in order to avoid boiling
overflow, cause an accident. Should be shut off the power, after
being pressure gauge pointer back to "0", and open the container
cover.
7. sterilizer after the termination of daily use, strike the
sewage drainage net within the container, the container and erase
scale, improve the quality of disinfection sterilization, as well
as energy saving and service life. If the container has more wipe
not to scale, can use the following solution cleaning scale. With
herself for: 10 kg of water, 1 kg caustic soda, 0.25 kg of kerosene
in container, soak about 1 hour, then wash scale, the final rinse
off with clear water.
8. sealing washer will use day long aging, affect the sealing
effect, should replace the sealing ring at this time.
9. Pressure gauge test at least once a year, if the pressure
gauge pointer back to "0", should be professional maintenance or
update, otherwise it will affect the disinfection effect of the use
of safe.