Disposable 72hours Child Type Closed Suction Catheter System For
Single Use
Feature
- Continuous Breathing
Simultaneous ventilation
- Soft Blue Suction Tip
This design reduces damage to the patient while suctioning.
After returning the catheter, the tip of the tube is easier for
irrigational cleaning.
- Disconnecting Wedge
Designed to equip the wedge with both disconnecting and clipping
functions
- For Endotracheal Tube and Tracheostomy Tube
Designed for endotracheal and tracheostomy tube using
Tubes with different length are available
- Prevent Cross Infection
Designed with a protective sleeve to isolate the bacteria inside
and help caregivers avoid cross-infection.
Packaging & Delivery
-Packaging Details
-Packing: 1pc/sterilized pouch, 10pcs/inner box, Outer packing:
100pcs/shipping carton
-Delivery Time: Within 20 days. It depends on the quantity of
order.
-Instructions for Use
*Set up Procedure
1. Inspect the product before use. Do not use it if the package is
not intact.
2. Open the sealed package and remove the product.
3. Connect the Endotracheal Tube/Tracheostomy Tube to the
Revolvable adapter.
4. Connect ventilator tube to the Revolvable Ventilator Connector.
5. Attach the date label to the color ring.
6. Before suctioning commences, ensure that the cap of the MDI Port
and Irrigation Port is closed.
7. Before suctioning: please ensure that the on-off valve is in the
open position. Simply slide the on-off valve to a position which
allows the catheter to enter the Endotracheal tube/Tracheostomy
tube.
*Suctioning Procedure
Caution-always use the recommended vacuum levels./Consider length
of suctioning time
1. Grip the three-way adaptor in one hand and use the other hand to
feed the suction catheter into the Endotracheal tube/Tracheostomy
tube to the required depth.
Depth markers can be visible through the protective cover for your
guidance.
2. Once the suction catheter is in the desired position /depth
depress the vacuum control valve to begin suctioning.
3. Remove the suction catheter until the protective sleeve is
straight.
4. Repeat steps1-3 as necessary.