Leeb231 1250µm Metal Shell Coating Thickness Gauge For Eddy Current

 

High Precision Changeable NFe Probe Metal Shell Coating Thickness Gauge For Eddy Current

 

 

Functions & Features

As the hot-sale item with  Changeable NFe probes and back-light screen.

This equipment(Leeb231) has two softward for PC connection and data analysis
User can switch off automatically or manually, and set the limit according to the different request.

 

 

Technical Parameter

Technical Parameters
Model No.Leeb231
Measuring principleNFe
Measuring range (µm)0~1250μm
ProbeChangeable
ShellMetal
Accuracy±(1~3%H+1) μm; H refers to the thickness of testing piece
Minimum resolution (µm)0.1μm
Min curvature of the min area (mm)Convex1.5 Concave9
Diameter of the min area (mm)Φ7
Critical thickness of substrate (mm)0.5
Memory300 groups measured data
Dimensions115*70*31mm
Power supplyAAA Alkaline battery
Standard ConfigurationMain unit, 1 probe & substrate,

5 calibration specimens (48.5μm,99.8μm,249μm,513μm,1024μm)

Optional AccessoriesProbes,Specimens

 

 
Basic Calibration Correction
     Basic calibration in the following cases, the change is necessary:
     -The probe frayed at the top.
     -Replace a new probe.
     -Special purpose.
     In measuring, if error is clearly beyond the scope, should be made for the
     properties of the probe to the calibration is called basic calibration.
     By entering the six calibration values (a thickness of zero and five values), can realign the measuring head.
     
 
Steps of basic calibration;
Step 1: Under normal condition, continuous press “Clear” three times;
Step 2: Press“Menu”and up & down key to select“Calibrate”,press“Menu”to into the correct Zero;
Step 3: Place the probe vertically in the middle of the matrix and hold still; after completion of shows such as 5 seconds away again.To the next.
Step 4: When show the“S1”(The first specimens),put the minimum value (48.5 μm)specimens on the standard substrate,probe vertically on the S1 hold still,
such as display again after 5 seconds.Display real thickness“Real thickness(0049.4μm)”,press“Menu”and up & down key to select to enter the value
from“(0049.4 μm)”to the real one“(0048.5 μm)” and go to the next;
Step 5: Display“S2”,remove“S1”,put ( 99.8μm)piece on the standard
substrate,probe vertically on the S2 hold still, such as display again after 5 seconds,display“real thickness(0096.7μm)”,press“Menu”and up & down key to select to enter the value from“(0096.7 μm)”to the real one“(0099.8um)”and go to the next;
Step 6: Use“Step 5”to complete the real value calibration of“S3(249μm),S4(513μm),S5(1024μm)”
Step 7: After calibrate S5will display“Loading...”,about 2 seconds later it will come back to boot interface.Measured again after 5 seconds, such as standard substrate(Or you will enter the continuous mode),correct Zero to ensure that the display value to 0 after the test piece (or measured values are not accurate). The basic calibration complete.
 
        Note:Specimens with other basic calibration, on the specimens of each thickness should be at least 1.6 times the thickness of a specimen.Such as:20,50, 100,250,500μm.
        After the calibration, the instrument will work according to the new calibration value.
 
 

High quality Equipment Box:

 

Hot Products

Changeable NFe Probe Film Thickness Gauge Eddy current Principle High Precision Changeable NFe Probe ...
Plastic Shell Magnetic Coating Thickness Gauge With Changeable Ferrous Probe Competitive Coating ...
Integrated Fe Probe Ferrous Metal Thickness Tester With One Specimen Calibration New system ...
Leeb231 1250µm Metal Shell Coating Thickness Gauge For Eddy Current High Precision Changeable NFe ...
FCC Car Paint Thickness Gauge Integrated Coating Tester With AAA Battery High Accuarcy Easy And ...
Leeb232 0.1μM Resolution Coating Thickness Gauge Changeable Fe NFe Probe High Precision Changeable ...