Introduction:

  • Three Phase Secondary Current Injection Test Kit s applied to test the relay unit in relay protection system. At the same time it could be considered as a universal three phase AC/DC voltage source or current source.
  • It is applied to test protection replay based on voltage and current operation.
  • The main function of the tester include: calibration of voltage and current operation relays, secondary circuit inspection,trip value of circuit breaker test, switch on and off time measurement of circuit breaker and so on.
  • Three Phase Secondary Current Injection Test Kit is a low cost, simple operation and reliable performance device for secondary circuit test.

 

 

Features:

  1. Intelligent mainframe that is used high speed DSP
  2. Separately running
  3. Microcomputer control, it can also connect with computer for operation
  4. 16 bit DAC chip
  5. Large LED screen display
  6. New type power amplifier
  7. Current, voltage output directly
  8. Self-protection
  9. Easy operation
  10. Abundant contact
  11. Compounded structure
  12. Abundant capabilities against low price

 

 

Specifications:

No.ItemsSpecifications
Alternating Current(AC) output
1Output accuracy0.5magnitude
2Phase current output (effective value)0~30A
3Three parallel phase current output (effective value)0~90A
4Phase current value with long time under permission (effective value)10A
5Phase current maximum power output420VA
6Three parallel phase current maximum power output900VA
7Three parallel phase current maximum work time10s
8Frequency range (base wave)20~1000Hz
9Harmonic frequency number1~20 time
Direct Current(DC) output
1Output accuracy0.5magnitude
2Current output

0 ~ ± 10 A/each phase

0 ~ ± 30 A/every three parallel

3Maximum output of load voltage20v
Alternating Voltage output
1Output accuracy0.5magnitude
2Phase voltage output (RMS)0 ~ 120v
3Line voltage output (RMS)0 ~ 240v
4Phase voltage/line voltage power output80VA/100VA
5Frequency range (base wave)20 ~ 1000Hz
6Harmonic frequency number1 ~ 20 times
Direct voltage output
1Output accuracy0.5magnitude
2Phase voltage output amplitude0 ~ +160v
3Line voltage output amplitude0 ~ + 320v
4Phase voltage/line voltage power output70VA/140VA
 
1AC current and voltage error<0.5%RDG+0.05%FS
2DC current error<0.5%RDG+0.05%FS
3DC voltage error1%RDG+0.1%FS
4Binary inputActive node 0~250V, 0~6V for “0”, 6~250V for “1”, passive node: 20mA max
5Binary outputDC 220V/0.2A, AC 220V/0.5A
The switch quantity and time measurement
1Switch parameters input

7 ways; Empty contact: 1~20mA, 24v;

Power contact access: “0”: 0-+6V;

“1”:+11 V-+250V;

2Switch parameters output2 couples DC: 220V/0.2A; AC: 220V/0.5A
3Time measurementMeasuring range: 0.1ms - 9999s / 0.1ms
 
1Power supplyAC220V ± 10%, 50Hz/60Hz
2Environment temperature0~50°C
3Humidity≤85%RH
4Weight<18KG

 

Hot Products

Product Introduction The smoke toxicity test device (squirrel cage method) is applicable to the GB/T ...
Roofing Fire Test Equipment with Burning Brands by ENV 1187 Test 1 and ISO 12468-1 Product ...
Product Introduction Smoldering characteristic test furnace is mainly used in the fire performance ...
Product introduction The thermal radiation flame propagation test equipment is developed according ...
Product Introduction The thermal radiation melt drop tester is developed according to the French ...
Introduction NF P 92-504 for the determination of horizontal flame spread rate of materials. This ...