Home Companies Advanced Instruments Co.,Limited

Proof and Comparative Tracking Index Testing Equipment by UL 746A

Advanced Instruments Co.,Limited

Contact Us

[China] country

Trade Verify

Address: Hi-tech Developing Industrial Area, Liuyang City,Changsha,Hunan Province,China

Contact name:Sandy

Inquir Now

Advanced Instruments Co.,Limited

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

Proof and Comparative Tracking Index Testing Equipment by UL 746A

Country/Region china
City & Province changsha hunan
Categories Switching Power Supply
InquireNow

Product Details

Proof and Comparative Tracking Index Testing Equipment

Introduction:

 

The tracking indices test equipment is designed according to GB4207 and IEC60112. It is suitable for lighting, low voltage electrical equipment, household appliances, electrical machines, motors, power tools, electronic equipment , electrical instruments. It also applies for insulation, plastic, electrical connectors, accessories industry.

Method for the determination of the proof and the comparative tracking indices of solid insulating materials . It is simple, accurate, reliable, and practical features.

 

 

Standard:

  • ASTM D 3638-12:Standard Test Method for Comparative Tracking Index of Electrical Insulating Materials
  • IEC 60112:test for thedetermination of the proof and comparative tracking indices of solid insulating materials on pieces taken from parts of equipment and on plaques of materialusing alternating voltages.
  • UL746A:(CTI) Comparative Tracking Index or Insulation Tracking Electrical Test

 

 

Feature:

 

1.The combustion chamber paint treatment; nice, corrosion-resistant; dimensions: 1200mm X 600mmx 1200mm;

2. All sample holders are made of stainless steel;

3. The Proof and Comparative Tracking Index Testing Equipment consists of voltmeter, ammeter, time overcurrent relays, power supplies, circuit control system, push button switches, lights and other components;

4. Automatic drip device includes dropping tube, drip nozzle, counting means counting device drip, drip control device, dropping the volume measuring tools;

5. Drip nozzle device can be adjusted; adjust the height of 30-40mm; ensure dripping position in the middle of the two electrodes;

6. The electrode means comprises platinum electrode, the electrode fixtures, 4mm gauge, etc., the electrode can be adjusted, removable;

7. Electrode width 5mm, thickness 2mm, 99.9% purity platinum, 30 to 60 degrees tip;

8. The electrode sample pressure adjustable, adjustable range (1.00N ± 0.001N); electrode distance 4.0mm ± 0.01mm, the angle 60 ° ± 5 °;

9. Proof and Comparative Tracking Index Testing Equipment is supplied with the sample stage;

10.The glass: Thickness ≥4mm;

11. The Comparative Tracking Index Testing Equipment is supplied with electrolyte: sodium chloride and alkyl naphthalene sulfonic acid reagent each bottle.

 

 

Safety control system:

1.The test voltage: 0 ~ 1000V adjustable

The maximum error of voltmeter displays : 1%;

2. Short-circuit current: (1 Guests 0. 1) A,

The voltage drop value does not exceed 10%

3. The test current exceeds 0.5A duration of more than two seconds, the system automatically shut down equipment;

4.Security door control system; the door opens, the test system is not working;

5.The electrode: 5mm × 2mm rectangular platinum electrode and brass electrode of each pair meet size requirements of IEC60112 Figure B.1;

6.The electrode: the angle formed between the electrodes is 60 ° disabilities 5 °, a pitch of (4 Soil 0.1mm);

7.The electrode: the sample pressure: 1 00N disabilities 0.05N;.

8.Drip system: (30 ± 5) seconds (on time 29.8S + open drip drip duration 0.2S) automatic counting, digital display (can be preset), 50 drops of time: (24.5 Soil 2) min;

9.Drip systems: a drop of fluid needle tip to the sample surface height: 35mm disabilities 5mm (attach a gauge for measuring your reference);

10.Drip systems: drip Weight: 20 drops: 0.380g ~ 0.489g; 50 drops:.. 0 997g ~ 1 147g;

11.PLC with touch screen control system; to achieve test smarter, more convenient has been operated.

12.The test data is automatically recorded, automatic storage, free retrieval;

 

 

Specification:

 

1Electrode materialelectrode tip - platinum metal (platinum)
2Electrode pressure1.0 ± 0.05N
3Electrode distance4.0mm ± 0.01mm, the angle 60 ° ± 5
4Electrode voltage100 ~ 600V (48 ~ 60HZ) adjustable, short-circuit current 1.0 ± 0.1A of less than 10%
5Dropping height30 ~ 40mm adjustable
6Dropping size44 ~ 55 drops / 1cm3 adjustable
7Dropping time30s ± 5s,
8Dropping times0 ~ 9999 times available
9Leakage judgmentwhen loop current is greater than 0.5A and maintaining for 2 seconds,the relay cut off the current and result indicates sample failed
10Testing processautomatic control, independent ventilation
11Applicatable standardIEC60112, ASTM D3638-12,UL746A
12Inside Volume0.1 m³ (Can be customized 0.5 m³ or 0.75 m³)
13Overal dimension630mm width x390mm depth x750mm height

 

Hot Products

Lateral Flame Spread Testing For Walls Surfaces Introduction Lateral Flame Spread Testing Equipment ...
Fire Propagation Index Testing Equipment for Laminated Foil BS476-6 FPT.pdf Introduction: BS 476-6 ...
ISO9239 FRPT.pdf Introduction: Critical Radiant Flux is the main parameters of the fire risk for ...
Introduction This Fire Test Equipment is developed base on the standard AS1530.2, test for ...
Introduction: The BS 476-12 Ignitability Tester is designed according to the BS476-12:1991. Method ...
Product Introduction The smoke toxicity test device (squirrel cage method) is applicable to the GB/T ...