ISO 6886 Oil Test Equipment for Vegetable Fats Oxidative Stability Rancimat Test

Country/Region china
City & Province changsha hunan
InquireNow

Product Details

Animal and Vegetable Fats and Oil

Determination Instrument of Oxidative Stability

OST-R3 Oxidation Stability Tester.pdf

Introduction:

This newly researched and produced instrument can determine oxidation stability of grease fastly and correctly.It provides effective scientific basis for researching the processing,using,storage,oil selection of vegetable oil production and usage of antioxidant and reinforcers.And it can also help for reseaching oxidation stability of environmental automobil fuel oil FAME.

 

It is mainly suitable to:

• determine animal and vegetable Fats and Oil oxidation stability
• evaluate anti-oxidation degree of antioxidant

• measure oxidation stability of cosmetics with rich grease

Related standards:

• ISO 6886:Animal and vegetable fats and oils -- Determination of oxidative stability (accelerated oxidation test)

• EN 14112-2003 :Fatty Acid Methyl Esters (FAME) - Determination of oxidation stability (accelerated oxidation test)
• EN15751 Automotive fuels - Fatty acid methyl ester (FAME) fuel and blends with diesel fuel - Determination of oxidation stability by accelerated oxidation method

• EN16568 Automotive fuels - Blends of Fatty acid methyl ester (FAME) with diesel fuel - Determination of oxidation stability by rapidly accelerated oxidation method at 120 Degrees

 


Features:

1.The softare records variation cureve of conductivity autoamtically,and judge induction inflection point by calculate second derivative of curve;

2.When the test is finished,the induction response time and curve will be stored data base for query.

3.Besides of induction response time,software can also evaluate the stable time,that is time for conductivity reach to certain value.

4.Software can re-evaluate every test curve.Through setting time range,select certain section of curve to evaluate,in order to avoid abnormal data caused by cleaning,sample or others.

5.Data base can store all related information,including conductivity curve,induationi response time,using method,start and ending time of test.

6.Users can set query terms to find test data;

7.Software also has functions of multi-curves comparison and analysis

8.Display mode:curve and digital

9.The temperature and flow can be calibrate by software.

10.Working units:4 tubes

11.Each sample is controlled independantly,and it can be set automatically according to requirement.

0~800 μS/cm

 

Specifications:

Temp.controllingImport PID temp.controller
Flow range8~12L/h±10%
Measuring range0~800 μS/cm
Resolution0.1μS/cm
Gas source10L/H with special membrane pump
Temp.controlling methodelectrical heating rod
Working temp.50~150℃±0.1℃
Power supplyAC220V±10% 50HZ
Power consumption1.1KW
Result processingautomatical calculation and storing by PC

 

Hot Products

ISO9239 FRPT.pdf Introduction: Critical Radiant Flux is the main parameters of the fire risk for ...
Introduction This Fire Test Equipment is developed base on the standard AS1530.2, test for ...
Introduction: The BS 476-12 Ignitability Tester is designed according to the BS476-12:1991. Method ...
Product Introduction The smoke toxicity test device (squirrel cage method) is applicable to the GB/T ...
Roofing Fire Test Equipment with Burning Brands by ENV 1187 Test 1 and ISO 12468-1 Product ...
Product Introduction Smoldering characteristic test furnace is mainly used in the fire performance ...