This instrument is designed and manufactured as per standards GB/T
261-2008 Determination of Flash Point - Pensky-Martens closed cup
method and ASTM D93 Standard Test Methods for Flash Point by
Pensky-Martens Closed Cup Tester. It is used to make determination
of the closed cup flash point of the petroleum products.
I. Main technical features
1. Single chip microcomputer control technique.LCD screen displays.
English man-machine interaction interface. The heating power is
continuously step-less adjustable. The rate of heating up is
exact.Easy setting for parameters.
2. The LCD screen has prompting menu,prompting input operation
interface. It shows the setted parameters, real-time temperatures
and other test parameters. Press the “record”button when the flash
point appears. The screen will show and reserve the flash point.
3. The structure is newly design and small.Stainless steel
table-board looks clean and beautiful. It can be used as long as
equipped with gas or other civilian fuels. It is easy to use and
test results exact.
II. Main technical specifications
Power supply | AC( 220±10% )V, 50Hz |
Total power consumption | ≤650W |
Heating device | (1) The furnace is silicon carbide material. Power is 600W. (2) The heating power is adjustable from 0 W to 600W |
Heating rate | Procedure A: (5~6)℃/min, Procedure B:(1~1.5)℃/min. |
Stirring device | (1)Stirring motor: BYGH101 stepping motor (2) Driving mode: flexible shaft (3) Shaft size: 8mm×40mm
|
Stirring rate | Procedure A: (90~120)RPM, Procedure B: (250±10)RPM |
Stand oil cup | (1)Internal diameter:50.7mm~50.8mm (2) Depth: 55.7mm~56.0mm (3) Marking depth of oil testing capacity: 33.9mm~34.3mm; (4) Oil test capacity: about 70ml;
|
Igniting device | (1) Igniting source: gas(or other civilian fuels,the same below) |
Suitable environment | Ambient temperature:(15~35)℃ Relative humidity:≤85% |
Dimension | 340mm×330mm×380mm Net weight:9kg |