Home Companies DONGGUAN YUYANG INSTRUMENT CO., LTD

IEC60331 Wire Testing Equipment Cable Fire Resistance Test For Circuit Integrity Under Fire Conditions

DONGGUAN YUYANG INSTRUMENT CO., LTD

Contact Us

[China] country

Trade Verify

Address: B1505, FORTUNE PLAZA, No.1, Yuanmei Road, Nancheng District, Dongguan City, Guangdong Province, China

Contact name:Seline Yu

Inquir Now

DONGGUAN YUYANG INSTRUMENT CO., LTD

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

IEC60331 Wire Testing Equipment Cable Fire Resistance Test For Circuit Integrity Under Fire Conditions

Country/Region china
City & Province dongguan guangdong
Categories Refrigeration & Heat Exchange Parts
InquireNow

Product Details

Cable Fire Resistance Test for Circuit Integrity Under Fire Conditions 

 

 

Introduction

This fire resistance test apparatus to be used for testing cables required to maintain circuit integrity when subject to fire alone or fire with shock where the test condition is based upon flame with a controlled heat output corresponding to a temperature of at least 750˚C or fire with shock at a temperature of at least 830 °C.

  • ​IEC 60331 Test for electric cables under fir conditions Circuit integrity

  • IEC 60331 Part 11 Fire alone at a flame temperature of at least 750˚C

  • ​IEC 60331 Part 21 Cables of rated voltage up to and including 0.6/1kV

  • IEC 60331 Part 12 Fire with shock at a flame temperature of at least 830˚C

  • ​IEC 60331 Part 31Fire with shock Cables of rated voltage up to and including 0.6/1kV

 

Standards:

  • IEC 60331-11&12:2002

  • IEC 60331-21~23:1999

 

 

Features

  • Heat Source : AGF Ribbon Type Propane Gas Burner

  • ​Structured with 3 rows of cross holes, Average front width of burner: 15mm, Hole diameter: 1.32mm, Distance between the center: 3.2mm.

  • ​A polished hole row in pilot hole for flame preservation in both faces of Burner plate.

  • ​Two thermocouples are horizontally installed separated 45mm left and right of burner center-line to check temperature.

  • ​Mass flow controllers to regulate the gas and air flow, the accuracy is not less than 1.5%.

  • ​Software can show the burner calibration temperature, propane and air flow rate etc.

  • ​The current is provided by a three-phase star transformer with sufficient capacity to maintain the required test voltage when circuit integrity cannot be maintained.

  • ​Conductor connection indicator (LED indicator) to form a current.

  • ​The voltage value is automatically set by software and adjusted by a panel controller.

  • ​Digital display of the current and voltage data, accuracy of current is 0.001A.

  • ​Current protection using fuse with 2A.

  • The real-time measurement data of all current and voltage are collected by computer, and the curves of current and voltage are generated.

  • ​When circuit integrity cannot be maintained, the corresponding indicator lamp is extinguished, the computer automatically determines the state and records the time, and sends out the sound and light alarm.

 

Installation requirements of Fire Resistance Test for Circuit Integrity Under Fire Conditiions

Electrical

AC 380V 3-phase, 50/60Hz, 30A

​Ambient Temperature

Operating 10°C to 35°C

Size

1750 mm (W) x 680 mm (D) x 1400 mm (H)

Weight

160KG

Gas

air and propane

 

Hot Products

Please email to Sandy@yuyangtest.com if you need details. Solar Cell Spread of Flame & Burning Brand ...
ISO 8301, EN 12667 Thermal Conductivity Tester of Heat Flow Meter 0.1 ~ 8.0 m2K/W Description: It is ...
100~450℃ Melt Flow Index Tester MFR MVR Thermoplastics ISO 1133 ASTM D1238 Description: The melt ...
BS EN136-1998 Medical Surgical Mask Flame Retardant Tester (800 ± 50) ℃ Description: The Flammabilit...
100~450℃ Melt Flow Index Tester MFR MVR Thermoplastics ISO 1133 ASTM D1238 Similarities ISO 1133 and ...
100~450℃ Melt Flow Index Tester MFR MVR Thermoplastics ISO 1133 ASTM D1238 ISO 1133 Vs. ASTM D1238 ...