Home Companies DONGGUAN DAXIAN INSTRUMENT EQUIPMENT CO.,LTD

YY/T 2626-2019 Mask Tester Area 100cm² With Stainless Steel Clamps

DONGGUAN DAXIAN INSTRUMENT EQUIPMENT CO.,LTD

Contact Us

[China] country

Trade Verify

Address: No.168 Zhen'an West Road Changlian Industry Mansion Xiabian Community Dongguan China

Contact name:Sophia Su

Inquir Now

DONGGUAN DAXIAN INSTRUMENT EQUIPMENT CO.,LTD

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

YY/T 2626-2019 Mask Tester Area 100cm² With Stainless Steel Clamps

Country/Region china
City & Province dongguan guangdong
InquireNow

Product Details

Particle filtration efficiency tester(PFE)
 
Scope of application:
Especially used for calibration: infrared forehead temperature gun, infrared ear thermometer, portable infrared thermometer medical thermometer, infrared body surface screening instrument;
 
Standards: 
GB / T 32610-2016 ,GB / T19083-2010 ,YY/T 2626-2019 ,YY0469-2011
 
Constituent part :
Aerosol generator
- It is mainly used for the test of high concentration aerosol,PSL standard particles of 0.3um are used, and the particle size range of the main particles emitted is ≥0.3um, accounting for more than 70% of the total, which can continuously and stably produce aerosols. Upstream of the filter material, the test particles are provided.
Particle counter
- It is mainly used to monitor the concentration of upstream and downstream particles of the measured filter material and obtain the efficiency of the measured filter material.
Differential pressure transducer
- It is mainly used to test the pressure difference (or resistance) of the measured filter material under a certain test flow rate.
Stainless Steel Clamps
- Mainly used for the pressure and sealing of the measured filter material.
Air pump
-Mainly used for the test flow through the filter material, stable flow, green energy saving, small vibration noise.
Electrical control and software
Circuit control and operation of equipment, calculation and data processing of software, stability judgment of upstream concentration, operation control, parameter setting. 
 
Technical Parameters:
Flow range
(20 ~ 100) L / min, accuracy 2%
Filtration efficiency
(0.001 ~ 99.999)%
Resistance
0 ~ 2000Pa, accuracy 2%
Test area
100cm²
Aerosol concentration
(12 ~ 20) mg / m³ (salinity), (50 ~ 200) mg / m³ (oily)
Test efficiency range45%-99.9%@≥ 0.3μ m
Test differential pressure range0-50pa± 1% , 50-500pa± 1.5%
Test flow of measured filter material25-100L/min±2.5%(adjustable)
Test aerosol typesPSL
 
- Test mode: initial efficiency and loading efficiency
- Test temperature: (25 ± 5) ℃
- Relative humidity: (30 ± 10)%
- External dimensions: 950 × 600 × 600mm
- Gas source: flow 198L / min (0.6Mpa) at 550kPa
- Power source: AC220V 50Hz
- Weight: 100kg

Hot Products

MT914-2008 Conveyor Belt Roller Friction Test Machine_Rope Core Flame Retardant Belt Roller Friction ...
ISO 9151 Protective Clothing Heat Transmission Tester BS EN 367 Product Description: Heat transfer ...
600°F Flame- Retardant Protective Clothing Testing Equipment GB 8965.1.2009 Scope of application: ...