[China]
Trade Verify
Address: 25F, China Resources Building, Xiejiawan, Jiulongpo District,Chongqing,400050,China
Contact name:Ivy Zhang
Chongqing gold mechnical and electrical equipment Co., Ltd |
Verified Suppliers
|
|
Lab IEC 60112 PTI And CTI Tracking Index Tester
Introduction:
The tracking index tester is a special testing instrument developed, designed and manufactured in strict accordance with GB 4207, IEC 60112, JIS C 2134 and other standards. It is suitable for research, production and quality inspection departments of lighting equipment, low-voltage electrical appliances, household appliances, machine tool electrical appliances, motors, power tools, electronic instruments, electrical instruments, information technology equipment, and is also suitable for insulating materials, engineering plastics, electrical connectors, and accessories industries. It simulates the determination of the tracking index and the resistance tracking index under humid conditions, and has the characteristics of simplicity, accuracy, reliability and practicality.
Tracking test is a test method used to evaluate the insulation performance of electrical equipment, IEC 60112. This test is used to detect whether the insulation materials of electrical and electronic products are prone to leakage and arcing (i.e. tracking) in humid environmental conditions.
Test principle:
It is used to evaluate the tolerance of the surface of solid
insulating materials under the combined action of electric field
and contaminated medium, and to determine its comparative tracking
index (CTI) and proof tracking index (PTI);
Comparative Tracking Index (CTI): The highest voltage value that
the surface of the insulating material can withstand 50 drops of
electrolysis (0.1% ammonium chloride aqueous solution) without
forming a leakage trace, in V.
Proof Tracking Index (PTI): The voltage required for the material
surface to form a permanent conductive carbon path after
withstanding the action of 50 drops of electrolyte at a rate of one
drop per 30 seconds, expressed in V.
Sample requirements
1. Plates and sheets with a diameter of 30 mm or more, or length
and width greater than 30 mm; thickness > 3 mm, samples can be
stacked, and the surface is flat;
2. Testable voltage: up to 600 V.
Main feature:
1. The combustion box is painted, beautiful, and
corrosion-resistant; all sample racks are made of stainless steel.
2. The equipment consists of a voltmeter, an ammeter, a time-delay
overcurrent relay, a power supply, a circuit control system, a
button switch, a signal light, etc.
3. The automatic dripping device includes a dripping cylinder, a
dripping nozzle, a counting device, a dripping counting device, a
dripping control device, a dripping volume measuring tool, etc.
4. The drip nozzle device is adjustable; the height is adjusted to
30-40mm; ensure that the dripping position is between the two
electrodes.
5. The electrode device includes a platinum electrode, an electrode
fixing device, a 4mm gauge, etc. The electrode is adjustable and
detachable.
6. The electrode is 5mm wide and 2mm thick, 99.9% pure platinum,
and has a 30~60 degree tip.
7. The pressure of the electrode on the sample is adjustable, with
an adjustment range of (1.00N±0.001N); the electrode distance is
4.0mm ± 0.01mm, and the angle is 60°±5°.
8. Equipped with a sample table.
9. Equipped with a glass plate: thickness ≥4mm.
Specification:
Electrode material | Electrode head - platinum (platinum) |
Electrode pressure | 1.0±0.05N |
Electrode distance | 4.0mm ± 0.01mm, angle 60°±5 |
Electrode voltage | 100~600V (48~60Hz) adjustable, voltage drop is less than 10% when short-circuit current is 1.0±0.1A |
Drop height | 30~40mm adjustable |
Drop size | 44~55 drops/1cm3 adjustable |
Drop time | 30s±5s |
Drop times | 0~999 times adjustable |
Leakage judgment | If the loop current is greater than 0.5A and remains for 2 seconds, the relay will cut off the current, indicating that the sample is unqualified |
Test process | The test procedure is automatically controlled, and the ventilation is independent; |
Reference standards | IEC 60112, GB/T 4207, GB 4706.1 |
Equipment dimensions | 1200mm wide × 600mm deep × 1300mm high |