Cheapest Factory Prices Disposable Nitrile Surgical Gloves in China

 

 

The requirements for medical-grade gloves are as follows
1:length

Measure with a ruler in mm, and align the glove naturally perpendicular from the edge of the glove to the zero degree of the

ruler, perpendicular to the middle finger tip of the glove.

 

 

 

2: thickness

Cuff: measure the upper lip of the roll 13+3mm at any of three different positions, take the average
Fingertip: cut from the middle fingertip down 13+3mm into four equal parts, and take the average of four parts.
Finger seam: cut the finger seam between middle finger and index finger, measure the thickest part on both sides and record the
maximum value.
3The leftover side slices are required to be placed in the center of the glove at a distance of 1cm from the edge of the windle,
and the width of the parallel part of the cutter center is 6mm. When the tension is made, the upper and lower clamps are set at a
distance of 40mm, and the test position is the center and 1cm above and below, and the average value is taken.
4:Hanging water testThe water requirement is 1000+50ML, the height of gloves is 40mm, the hoisting time is 3 minutes, and the water temperature is
required to be 15-30°
5:Visual Inspecting TestHole/scratch/tear/drop of oil/mixed number/bad side/insect/hair/serious deformation = unqualified product, not available for sale.
6:Physical test frequencyNormally, the weight/length is measured once per hour, the thickness is measured once for 4 hours, and the tension/elongation is
measured once for 2 hours.
7:Standard for grade pinhole testingEach time each line each side sampling 75, 0-1 pinholes for Best quality grade. Main requirements: test all gloves on the production
line, test and wear them every 2-3 hours, no less than 10 minutes each time, and make relevant records.

 

Hot Products

Low MOQ Made In China Consumable Non Woven 3 Ply protective Disposable Face Mask Product Name 3 ...
CE Premium 50pcs per box Packaging 3ply Disposable Earloop Medical Surgical Face Mask with 3 ply Non ...
Disposable 3Ply 3 Ply Non Woven Dust Mouth Mask Medical Dental Doctor Surgery Surgical Face Masks ...
Surgical Medical Procedure 3 ply Face Mask Disposable Medical Mask Virus PM2.5 Mask Material 3ply ...
China Manufacturer 3 ply Disposable Nonwoven Surgical Face Mask Medical Face Mask Supplier 1. Single ...
3 Ply Virus Filter Protective Health Medical Face Masks Disposable Surgical Mask 3ply Medical Mask ...