Medical Face Mask BFE Testing Machine 2200cfu 28.3 L/Min

Country/Region china
City & Province suzhou jiangsu
Categories Measuring & Analysing Instrument Processing Services
InquireNow

Product Details

 

SJ-X100 Medical Face Mask Filter Tester BFE Testing Machine

 

Face Masks bacterial filtration efficiency BFE tester is to determine the bacterial filtration efficiency (BFE) of medical face mask materials by employing a ratio of the upstream bacterial challenge to downstream residual concentration of the tested medical face mask materials.

 

Features of BFE filter tester

  • Safe working environment, use professional negative pressure biological cabinet, ensure the safety of operators;
  • High negative pressure working cabin, two-stage high efficiency filter, 100% safe discharge
  • Use two-way six-level Anderson sampling;
  • Built-in peristaltic pump with adjustable flow rate;
  • Special microbial aerosol generator, adjustable spray flow of bacterial liquid, good atomization effect;
  • Industrial grade 10.4-inch color touch screen control, make the operation is simpler;
  • Microcomputer is used for control and data processing, which can be seamlessly connected to the laboratory information management system;
  • USB interface, support data transfer;
  • The safety cabinet has built-in LED lighting for easy observation;
  • Built-in ultraviolet UV disinfection lamp;
  • The front-opening sealed glass door is convenient for operation and observation;
  • Use imported isolation operation gloves, safe and reliable;
  • The system has functions of user authority management and audit trail;
Main parameter of BFE filter testerRangeResolutionAccuracy
Anode quality control sampling flow28.3 L/min0.1 L/min±2%
Sampling flow28.3 L/min0.1 L/min±2%
Spray flow0.1 ~ 10 L/min0.1 L/min±5%
Peristaltic pump flow0.01~3 mL/min0.01 mL/min±2%
Operating temperature0~50 ℃
Biosafety zone negative pressure-90 ~ -120 Pa
Data storage capacity20G
Data output and interfaceUSB(3)
HEPA performance≥99.995%@0.3μm,≥99.9995%@0.12μm
Total number of positive quality control sampler particles2200±500 cfu
Aerosol generator mass median diameterAverage particle diameter (3.0±0.3 µm),geometric standard deviation≤1.5
Six-stage Anderson sampler capture particle sizeclassⅠ>7 µm; classⅡ(4.7~7 µm); class Ⅲ(3.3~4.7 µm); class Ⅳ (2.1~3.3 µm); classⅤ(1.1~2.1 µm); class Ⅵ(0.6~1.1 µm)
Aerosol chamber specifications600 x 80x 3.0mm (Length diameter thickness)
Ventilation flow of negative pressure cabinet>5m3/min
Host size

Working area: 870*600*1100mm

Dimensions:1250*700*2000mm

Working noise< 65db
Working powerAC220±10%,50Hz,1KW

 

Hot Products

Y09-350 High Quality Laser Dust Particle Counter Sensor The portable laser particle counter ...
Air Particle Counter 0.3 Sensitivity For Food Processing Area The Air particle Counter Meet the ...
Y09-3016hw Portable Laser Handheld Airborne Particle Counter Y09-3016HW Handheld Air Particle ...
Y09-3016HW Handheld Air Particle Counter For Pharma Cleanroom​ Y09-3016HW Handheld Particle Counter ...
Y09-3016HW Laser Dust Particle Counter (2.83L/Min) For Cleanroom Y09-3016HW laser dust particle ...
Y09-350 Portable 6-Channel Particle Size Airborne Particle Counter Y09-350 laser dust particle ...