Tumor Marker PGI Pepsinogen I for Fully Automatic Immunoassay Analyzer IVD Reagents 100T/box,24T/box

INTENDED USE

The kit has been designed for the quantitative determination of Pepsinogen I (PG I) in human serum.

The method can be used for samples over the range of 1.0-350.0 ng/mL.

The test has to be performed on the CIA fully auto analyzer (Including CIA 600, CIA 1200, CIA 1800, CIA2800).

 

KIT COMPONENTS

Material Supplies

Reagent Integral for 50 determinations
Magnetic beads: TRIS buffer, magnetic beads complex binding with goat anti-FITC, buffer with BSA1×1.5mL
Calibrator Low:BSA buffer with low level of PGI antigen1×0.15mL
Calibrator High:BSA buffer with high level of PGI antigen1×0.15mL
PGI Anti A: PGI Antibody complex labeled FITC, containing buffer with BSA1×3.0mL
PGI Anti B: PGI Antibody complex labeled by AP, containing buffer with BSA1×3.0mL
All reagents are provided ready-to-use.

 

Reagent Integral for 100 determinations
Magnetic beads: TRIS buffer, magnetic beads complex binding with goat anti-FITC, buffer with BSA1×3mL
Calibrator Low:BSA buffer with low level of PGI antigen1×0.15mL
Calibrator High:BSA buffer with high level of PGI antigen1×0.15mL
PGI Anti A: PGI Antibody complex labeled FITC, containing buffer with BSA1×6.0mL
PGI Anti B: PGI Antibody complex labeled by AP, containing buffer with BSA1×6.0mL
All reagents are provided ready-to-use.

 

Reagent Vials in kit box
Control 1: BSA buffer with low level of PGI Antigen1×0.15 mL
Control 2: BSA buffer with high level of PGI Antigen1×0.15 mL
target value refer to Quality Control sheet

 

Accessories Required But Not Provided

SubstrateREF: SR2001
System Washing solutionREF: WR2001
Probe Washing SolutionREF: WR2002

 

product compositionTwo points: LH
Six points: ABCDEF
Testing method1. Treat the calibrator as the sample to be tested, and place it in the sample loading area of the instrument after mixing. Note that when mixing Avoid air bubbles.
2. During the detection, the system detects the luminescence value (RLU value) of each calibration point, and calculates the luminescence value according to the nominal concentration and luminescence value.
The calibration curve is automatically fitted by the instrument. In order to ensure the reliability of the calibration curve, the determination of the calibrator should be
Make a double.
3. The generated calibration curve can be used for 28 days, but it should be re-calibrated after the following conditions:
① Change of batch number of supporting testing reagents;
Change of batch number of luminescent substrate
②The measured value of the quality control product is not within the scope of quality control or the test results are inconsistent with the clinical situation;
③System for major repairs or maintenance.
Interpretation of test results1. For reasons such as methodology or antibody specificity, the same sample was tested using reagents from different manufacturers.
Testing with different reagents may give different test results, therefore, the results obtained by testing with different reagents should not be directly
They are compared with each other in order to avoid false medical interpretations.
2. The quality control product can be used as a reference for the reliability of the current experimental results, and its measured value should be within the
within the range allowed by the quality control sheet. Test results should be based on reference value ranges and other clinical factors and results
Comprehensive judgment, when the test result is close to the upper limit or lower limit of the reference value range, it can be considered
samples for confirmation testing.
Storage and expiration dateStored at 2-8°C, the product is valid for 12 months. Reagents should be used within 1 month after opening, and stored at 2-8°C away from direct sunlight.

 

Hot Products

pet animal detection analyzer monotest single reagent strip With cat dog serum/heparin anticoagulate...
IVD Pet animal cat detection reagent single test monotest Total Thyroxine T4 In Thyroid Clinical ...
IVD Pet animal dog detection reagent single test monotest Total Thyroxine T4 In Thyroid Clinical ...
Tumor Marker HE4 Human Epididymis Protein 4 for Fully Automatic Immunoassay Analyzer IVD Reagents ...
Tumor Marker CYFRA 21 Cytokeratin 19 Fragment for Fully Automatic Immunoassay Analyzer IVD Reagents ...
Tumor Marker CA242 Carbohydrate Antigen 242 for Fully Automatic Immunoassay Analyzer IVD Reagents ...