Thermal Analysis System Carbon Silicon Analyser for Molten Iron
Main Features:
The analyzer adopts electromagnetic interference resistance, dust proof, ultra-thin and portable design, easy to operate, non-professional personnel can operate, can be connected with other electronic equipment.
The analyzer is equipped with multiple lines, in view of different type of molten iron and the actual situation of molten iron, each factory can choose the suitable line, to make the detection more scientific and accurate.
The analyzer works by improved evaluation methods, can automatically control the important metallurgical parameters, satisfy the requirement of the foundry production quality control; can remedy deficiency what "spectrum" is difficult to measure nonmetallic elements (C, Si) accurately, and the time requirement that conventional instrument can't satisfy the rapid analysis .
Main technical parameter:
8. Working parameter: Power AC220V/50HZ,30W,
Environment temperature 0~50℃
9.Temperature compensation range: 0~15℃
10.Matched Carbon Cup type: K
11.Temperature Unit: Fahrenheit or Celsius
12. Max measurement time: 240seconds( generally 1 minute 45 second)
13.Display of measuring condition: display circularly Green Light, Yellow Light, and Red Light for “prepare”, “measure” and “finish”.
14. Protection standards: IP65 protection standard
15. EMC Standard: Perform anti-interference standard of EN50081-2 and EN50082-2.
16. The instrument can directly display the test result,and print automatically CEL,C,SI,Strength of extension(Rm), Hardness(HB), Peak, TL,TS and SC.
Error Code and Definition
Error Code | Definition |
1 | Casting temperature is greater than maximum scale value |
2 | Thermocoupleis broken |
3 | Casting temperature is less than 1200℃ |
4 | The different value between the peak temperature and the liquidus temperature is less than 20℃ or 40℃. |
5 | Casting temperature is less than minimum scale value |
6 | Carbon content is greater than carbon equivalent. |
7 | △T is negative |
8 | Solidus temperature is less than 1050℃ |
9 | No solidus temperature |
10 | Beyond temperature range. |
Packing list:
Item | Qty | Remark |
Instrument | 1 | Standard power line (attached) |
Support | 1 | Compensating cable (attached) |
Manual | 1 |