Home Companies Changsha Panran Technology Co., Ltd.

Wide Range PR320A 40*600mm Thermocouple Furnace

Changsha Panran Technology Co., Ltd.
Active Member

Contact Us

[China] country

Address: Room 504,Floor 5th,Chuangye Building, Hu-Da park,No186,Guyuan Road,High-Tech District,Changsha,Hunan ,China. (Zipcode: 410205)

Contact name:PANRAN

Inquir Now

Changsha Panran Technology Co., Ltd.

Wide Range PR320A 40*600mm Thermocouple Furnace

Country/Region china
City & Province changsha hunan
InquireNow

Product Details

A Wide Temperature Range Calibration Furnace For Thermocouples

 

The Thermocouple Calibration Furnace is a horizontal, open-ended tube furnace with a temperature range of 300 °C to 1200 °C. It is used for comparison calibration of noble and base-metal thermocouples by secondary high-temperature labs and instrument shops in industries such as aerospace, automotive, energy, metals, and plastics.

 

The PR320 series is the most accurate, reliable, and flexible furnace in its class, meeting the demanding requirements of high-temperature thermocouple calibration

 

  • Products model selection table
No.NameModelTemperature rangeFurnace sizeDimension (mm)Net Weight (kg)Power (KW)Isothermal block
1Thermocouple calibration furnacePR320A300~1200℃Φ40*600700*370*45026.12.5optional
2Base metal thermocouple calibration furnacePR320B300~1200℃Φ60*60031.52.5/
3Sheathed thermocouple calibration furnacePR320C300~1200℃Φ40*60027.32.5PR1142A
4Thermocouple calibration furnacePR320D300~1300℃Φ40*60026.12.5optional
5Base metal thermocouple calibration furnacePR320E300~1200℃Φ40*60027.32.5PR1145A
6Short type Thermocouple calibration furnacePR321A300~1200℃Φ40*300310*255*290113.0Optional
7PR321CΦ16*30010.53.0/
8PR321EΦ40*30012.43.0PR1146A
9High temperature thermocouple calibration furnacePR322A300~1500℃Φ25*600620*330*460453.0/
10PR322B300~1600℃Φ25*600433.0/
11Thermocouple Annealing FurnacePR323300~1100℃Φ40*10001010*260*36029.42.5/

 

 

 

  • Detailed technical parameters of the products

 

1. PR320A/D Thermocouple Calibration Furnace

 

ModelPR320APR320D
Application

S,R standard thermocouple,working S,R precious and base-metal thermocouple calibration

(need to add PR1145A Isothermal block)

Temperature range300~1200℃300~1300℃
Furnace SizeΦ40mm*600mm
Temperature field centerDeviate from the geometric center does not exceed 10mm
Temperature field widthThe temperature difference within 80mm ≤1℃
Temperature field gradientAdd a coaxial cleaning porcelain tube of about 20mm shall be installed in the furnace,Not exceed 0.4℃/CM within ±20MM range of temperature field center

 

 

2. PR320B Base metal thermocouple calibration furnace

 

Applicationbase-metal thermocouple calibration
Temperature range300~1200℃
Furnace SizeΦ60mm*600mm
Temperature field centerDeviate from the geometric center does not exceed 10mm
Temperature field widthThe temperature difference within 60mm ≤1℃

 

 

3. PR320C Sheathed Thermocouples Calibration Furnace

 

 

Temperature range300~1200℃
Furnace SizeΦ40mm*600mm
Size of Isothermal blockSize Φ8mm*7pcs, depth 90mm
Temperature field indicators 1 (with Isothermal block)From the bottom of Isothermal hole, the internal temperature difference is ≤0.5 ℃ in the axial direction of 30mm, and the absolute value of the temperature difference between any holes in the same section is ≤0.25 ℃
Temperature field indicators 2 (without Isothermal block)The deviation between the center of the maximum uniform temperature field and the geometric center of the furnace along the axis is ≤10mm;Within the range of a uniform temperature field with a length of ≥60mm and a radius of 14mm, the temperature field between any two points is ≤1℃
Temperature field indicators 3 (with Isothermal block, use a coaxial cleaning porcelain tube of about 20mm)The maximum temperature point in the furnace is ≤20mm away from the geometric center of the furnace, and the maximum temperature ±20mm has a uniform temperature field with the temperature change gradient ≤0.4℃/10mm
Implementation standard,
Standard itemPR1142A Sheathed thermocouple calibration furnace Isothermal block
Optional itemCoaxial cleaning porcelain tube of Φ20mm

 

4. PR320E Thermocouple Calibration Furnace

 

 

Temperature range300~1200℃
Furnace SizeΦ40mm*600mm
Size of Isothermal blockExternal Φ38mm, innerΦ28mm, depth 150mm
Temperature field indicators 1 (with Isothermal block)Within the axial direction of the effective working area, the absolute value of the temperature difference between any two points is ≤0.5℃;Within the radial radius of ≥14mm, the absolute value of temperature difference between any two points on the same section is ≤0.25℃
Temperature field indicators 2 (without Isothermal block)The deviation between the center of the maximum uniform temperature field and the geometric center of the furnace along the axis is not exceed 10mm;Within the range of a uniform temperature field with a length of ≥60mm and a radius of 14mm, the temperature field between any two points is no more than 1℃
Temperature field indicators 3 (with Isothermal block, but add a coaxial cleaning porcelain tube of about 20mm)The maximum temperature point in the furnace is ≤20mm away from the geometric center of the furnace, and the maximum temperature ±20mm has a uniform temperature field with the temperature change gradient ≤0.4℃/10mm
Implementation standard,
Standard itemPR1145A Sheathed thermocouple calibration furnace Isothermal block
Optional itemCoaxial cleaning porcelain tube of Φ20mm

 

5. PR321A Short Thermocouple Calibration Furnace



ModelPR321APR321CPR321E
Temperature range300~1200℃
Furnace sizeΦ40mm*300mmΦ16mm*300mmΦ40mm*300mm
ApplicationShort type TCShort S,R typeShort base metal
Temperature field indicators

PR321A: The temperature difference within 40mm ≤1℃

PR321C: The maximum temperature deviation furnace axial geometric center shall ≤10mm, in the test deviation from the axial geometric center of 20mm, the axial temperature gradient shall ≤0.4℃/mm

PR321E: Including Isothermal block: the absolute value of the temperature difference between any two points shall ≤1℃,and the absolute value of the temperature difference between any two points on the same section of the hole bottom shall ≤0.5℃, calculated from the axial direction of the hole bottom within 30mm

Implementation standard JJG 668-1997Short base metal TC calibration

 

6. PR322 Series High-temperature Thermocouple Calibration Furnace

 

 

ModelPR322APR322B
Temperature range300~1500℃300~1600℃
Furnace sizeΦ25mm*600mmΦ25mm*600mm
Heating materialSilicon molybdenum rodPlatinum rhodium 30
Cold resistanceAbout 0.1ΩAbout 3.5Ω
Max current50A16A
Operating voltage20~50V50~180V
Control cabinetPR351PR352
Temperature field centerDeviate from the geometric center does not exceed 20mm
Temperature field indicatorsthe maximum temperature ±20mm has a uniform temperature field with the temperature change gradient ≤0.5℃/10mm

 

 

The PR322 series is equipped with a special power control cabinet:

1. Adopts patented multiple over-current protection, and is provided with power-on soft start, heating current limitation, freewheeling protection, automatic stop and other functions.

2. No manual voltage gear shift or meter adjustment is required for power-on and heating process.

3. Equipped with RS485 and RS232 dual-communication connections.

4. Configured with ZRJ series calibration system software, the functions of start/stop, real-time recording, parameter query setting, etc. can be achieved.

5. While protecting the safety of the equipment, the manual operation is greatly simplified.

 

7. PR323 Thermocouple Annealing Furnace

 

 

Temperature range300~1100℃
Furnace sizeΦ40mm*1000mm
Temperature filed centerOne end of the temperature field is <100mm from the furnace
Temperature field widthwithin 40mm ≤20℃

(Please note seal test or open test when ordering)

Hot Products

Standard Liquid Calibration Oil Bath Is 90°C To 300°C The PR500 series uses liquid as the working ...
Best Digital Temperature Control Available Temperature Calibration Baths Depending on the temperatur...
Low Temperature Refrigerated Temperature Calibration Baths The PR500 series uses liquid as the ...
PR532 Series Cooling Calibration Bath Feature 1. Temperature control system adopts PID temperature ...
Heat Tank 0~100℃ Liquid Calibration Bath Depending on the temperature range, PANRAN calibration ...
High Performance Temperature Sensor Calibration Furnace With Isothermal Block PANRAN TECHNOLOGY as a ...