Double Door Hospital Steam Sterilizers With Water Saving System

Country/Region china
City & Province jiangsu
Categories Rubber Processing Machinery

Product Details

                 

Double Door Hospital Steam Sterilizers With Water Saving System

 

Description

 

The double door Hospital sterilizer is compliant with CSSD standards as well as all other additional and relevant guidelines.

 

This particular device features vertical sliding glass doors.

 

A highly compact and very efficient machine, this device is useful in hospitals for the sterilization of rubber goods, textiles, glassware, containers and various instruments

 

Materials

• Material of Vessel, Jacket and Generator : AISI 304
• Exterior body : AISI 304
• stainless steel trolley for easy transporting

Programs

• 121 ºC Rubber
• 125 ºC Rubber
• 134 ºC Instrument
• 134 ºC Textile
• Bowie-Dick Test
• Leakage test
• Optional programs

Control Sysyem


• 5.7" colour touchscreen display
• Full automatic with microprocessor or PLC

 

Main Parameters

 

Model

Chamber Size

Overall Size

Net Weight

Steam

Water

Power

W*H*D mm

W*H*D mm

Kg

Kg/cycle

Kg/cycle

KVA

MD-0.30V

600*600*840

1200*2000*1180

850

22

35

2

MD-0.36V

600*600*1000

1200*2000*1340

900

25

40

2

MD-0.45V

600*600*1250

1200*2000*1590

980

27

50

3

MD-0.60V

600*600*1670

1200*2000*2010

1100

30

60

3

 

Custom Design for Limited Space

 

Medpharm’s strong engineering capability allows for highly customized autoclave design.

 

Facilities with space layout limitations will benefit from custom narrow design autoclaves

which provide high capacity in a limited space.

 

Narrow design autoclaves allow for more autoclaves in a given space to ensure continued sterile processing in case of autoclave downtime.

 

Stainless Steel Piping

 

Stainless steel piping and connectors are used to avoid corrosion which is necessary to safeguard the integrity of the autoclave and prevent contamination of the sterilized loads (either as chemical or particulate contamination).

 

 

Brochure load down from here

Medpharm medical large steam autoclave brochure.pdf

Hot Products

High Vacuum Degree Sterilizers Autoclave With Low Noise Water-ring Vacuum Pump For Silence ...
Liquid Cycle With Pressure Ballasting Steam Sterilization Using Autoclave Feature 1. Hinged door ...
High Pressure High Vacuum Autoclave Sterilizers Ensuring Perfect Sterilization For Hospital And ...
Rectangular Chamber With Smooth Round Corners And Jacket Medical Instrument Sterilizer Machine Model ...
Overpressure Protection Autoclave and Sterilizers With Safety Door System Main parameter PLC based ...