cavitation+Tripolar RF+ Monopolar RF for weight loss cellulite loss beauty equipment machine
Product introduction
1. Specification:
TFT LCD Display 8 inch true color touch screen
Technology Cavitation+(Monopolar RF+Tripolar RF)
Cavitation frequency 40KHz
Cavitation Max Power 80W
Tripolar RF 1MHz (output frequency);
Monopolar RF 1MHz (output frequency );
Handpieces 4 pieces
1).Monopolar RF head
2).Tripolar RF head (one big, one small)
3).Ultrasonic cavitation head
Voltage 220±10 V AC/110±10 V AC 50/60Hz
Size 49cm×31cm×25cm(L×W×H)
Weight 15KG
2. Product features
1.1.1 Body shape instrument pick the Israel hole blasting fat
technology, RF technology and vacuum melting fat pinch drainage
techniques in one set with multi-body sculpting, at the same time
have a 40K strong sound wave system, and the strong sound
waves can produce a liquid implosion effect, which is the wave of
expansion and compression then make the liquid in the formation of
a large number of micro-voids, these voids are gases and steam ,
With the compression cycle, the strong sound wave cause pressure
effects for liquid molecules, and in the expansion cycle, resulting
in negative pressure effect.
In fact, inside the liquid or biological tissues exist a cohesive,
in the low density of fat cells, molecular bonding will be weak,
and the strong sound waves can generate a low negative pressure
which can produce organizational gap that in physics known as "hole
phenomenon."
The direct affects of implosion which micro-gap inside and outside
the Cells results in enhancing the molecular motion, so as to
achieve a high energy level, eventually lead to the breakdown of
fat cells.
Mono-polar and tri-polar RF system, not only effect of the skin but
also the subcutaneous fat, so it can be widely applied to the face,
neck wrinkle removal and tighten, as well as other parts of the
body to lose weight, improve the orange peel-like skin , postpartum
abdominal tightening and so on.
It is the radio frequency heat energy through the epidermal and
stimulate the epidermal deep tissue, make the subcutaneous
tissue produce immediate tighten the collagen fibers and
regeneration. The skin becomes plump smooth again, restore the
original flexibility, while the heat effect, can enhance blood
circulation, improve microcirculation and inhibit sebaceous gland
produce, which results in eliminating wrinkles, shrinking pores,
tighten skin, removal Acne.
While the vacuum system was first used in medicine to be cured
scars, liposuction post-operative recovery, as the technology
development, and then someone found that the deep massage is
equivalent to a passive aerobic exercise, it has unique effect of
expulsion of toxin in the body and improve the orange peel tissue .
body sculpting instrument is a novel, non-surgical, non-invasive,
painless, safe and efficient beauty equipment.
1.1.2 The instrument is equipped with four hand-held probe, which
can be accordance to the different parts of the treatment needs to
select the proper size and treatment system.
1.1.3 This unit pick intelligentized control, touch screen
operation, visual interface application, which ensure the guest
easy to learn and safely use.
1.1.4 This unit use continuous output mode, which accordance with
acting on the human body parts and operational methods for
enhancing operation. Energy output is divided into 50 files, Ⅰ ~ 50
files of energy from low to high, according to the needs of
physical therapy choose to use a different stall; acting on the
body's energy value range is from 1J to 50J, when in operation,
users can accord to the patient’s capacity to adjust the energy,
increase or decrease in volume of 1J.The working hours are dynamic
countdown display; Each program is sorted, refined, and processed
by the highly experienced physicians, which based on several years
of clinical nursing experience, and combined with the scientific
theory of human life. And then be abstracted into a certain law of
mathematical functions, and process by computer.
1.1.5 For the convenience of users, we design interactive interface
with automatic countdown and remind function.
1.1.6 This unit use SCM control and LCD display,
The main components pick large scale integrated circuits,
clear vision and high reliability. The strong sound wave system
uses the best 40K ultrasonic frequency for body shaping, This
frequency can be effective for treat stubborn cellulite and orange
peel fat, which use sound waves of the cavity effect, acting on the
fat group and break them down and then absorbed by lymphatic
system. RF system uses 1MHz RF frequency, which can make cells
produced a strong resonance of water molecules rotate and produce
heat by the friction. Thereby heat up to the collagen and fat cells
for the purpose of heating. Vacuum system uses adjustable pulse
control, and thus can be able to mimic the effects of high-level
massage doctor
1.1.7 This unit use modular design and has clear structure ensure
the user easy to install and maintain.
3. Application:
cavitations:blast fat on arm, abdomen, buttocks and thigh , and
improve the skin.
6.3RF: cheek, lower jaw, neck
skin tightening; nasolabial fold improvement; waist and abdomen,
limbs, weight loss; hips, big legs, orange peel-like skin
improvement.
4. Taboo:
a. Pregnant and in the menstrual period women are prohibited use in
abdomen; such as fever, suffering from infectious and acute
disease; have heart disease, config heart pacemaker, severe
high blood pressure, cancer disease, asthma, deep vein thrombosis,
varicose veins , goitre, cancer, epilepsy and other patients; has
hemorrhagic disease, trauma, vascular rupture, skin inflammation,
skin disease patients and pregnant women; immune system
dysfunction, etc. over-sensitive to the current person can not use
this device
b. If feeling unwell, do not use this device. Prohibited in the
ears, nose, eye, throat and other parts used; when use the
cavitation must pay attention to: female uterus, breasts, genitals,
medical shaping the site, there are metal parts of the body and the
filler parts, can not use this device. If the skin feels any
discomfort or pain, do not use this device. Can not operate in the
back, we must avoid the bone bits.
c. The body buried in hearing aids, heart pacemakers, artificial
hearts
and other around medical electronic devices are banned on the use;
d. Children under 12 years old
1.3 Product structure
1.3.1 The appearance structure as the picture shows, This unit main
include: main unit, handle(contain Probes) and Cable.
Treatmenthe1 | Treatment head2 |
Power line | Treatment head3 |
1.3.2 The probe is divided into four kinds, see below table.
Table 1
Probe number | Probe size | Main acting parts |
1#cavitation | Φ78 | Limbs , abdomen and buttocks |
2#6RF(big) | Φ65 | bigger parts of body |
3#3RF(small) | Φ45 | small parts of body |
1.4 Main Specification
TFT LCD Display | 8 inch true color touch screen |
Technology | Cavitation+(Monopolar RF+Tripolar RF) |
Cavitation frequency | 40KHz |
Cavitation Max Power | 80W |
Tripolar RF | 1MHz (output frequency); |
Monopolar RF | 1MHz (output frequency ); |
Handpieces | 4 pieces 1).Monopolar RF head 2).Tripolar RF head (one big, one small) 3).Ultrasonic cavitation head
|
Voltage | 220±10 V AC/110±10 V AC 50/60Hz |
Size | 49cm×31cm×25cm(L×W×H) |
Weight | 15KG |
5.Install and use
2.1.1 Plug the power line into the power and unit
2.1.2 According to different needs, when replace the probes of
different sizes, hold down the interface to pull out the end of the
replacement, and then in accordance with the direction of the gap
can be connected.
2.1.3 when in Operation, the probe face must be close to the skin,
press the default settings of the parameters from the minimum to
begin, and then adjust based on individual level to gradually
increase the role of energy and time.
2.2 power on
For security, please turn the power orderly ; First, make sure you
provide 3-pin power outlet is good and reliable . Second, the
instrument's power plug and power outlet is a closely contact.
Finally, ensure the two prerequisite conditions are correct, please
turn on power switch behind the unit. After turn on the switch,
then turn on the power key. The host will be powered on and go into
the user interface.
After Power on, the boot screen will display the company LOGO ,
point at any icon into the language selection interface, such as
below graphics shows:
After choosing the language then into the main menu system interface, as below figure shows:
2.3 choose treatment system then into the operation interface shown in Figure
Adjust the parameters then to begin treatment.
2.3.1 Energy output is 50 files when in working state and
adjustable from low to high
6 Warning
3.1 Main unit
3.1.1 Please use the random allotment of three-core power cord and
prohibit using two-core power cord. middle of the core in Power
plug must be reliable to ground.
3.1.2 Inoperable areas: the eye, ear and other parts of taboo.
3.1.3 Avoid in strong magnetic fields and radio waves occurred in
the custody places, do not use in the bathroom, kitchen and other
places severe moisture and dust or high temperature storage areas.
3.1.4 Apart the host is strictly prohibited, in addition to the
company's authorized officer, others should not open the lid;
prohibit non-professional staff carry out repairs; otherwise own
risk.
3.1.5 Unused for more than six months and re-use have to accept
check-up services, in order to long-term use of this product,
please buy regular services in shop.
3.2 handle
3.2.1 When the hand-held pieces in the event of the failure or
damaged and need to replace, please unplug the hand-held pieces of
sockets. When install the handle please accordance to the direction
of the connector on the chassis, and inserted in the end force to
ensure that the probe connector with the socket are close. During
the installation process please avoid holding pieces fall on the
ground.
3.2.2 Dismantling or transformation of hand-held items is strictly
prohibited.
3.2.3 Do not strongly impact or percussion hand-held pieces.
3.2.4 when in operation, the probe should be completely
contact with the skin, and then begin to operate(start-point touch
screen).
3.2.5 If replace person after each treatment, the energy of the
interface must be returned to the lowest setting, and then
adjust the energy accord to each individual level and the
degree of sensitivity!
3.2.6 Others functions are prohibited except this user manual
specified..
3.2.7 Probes clear:Available medical alcohol scrub.
3.2.8 Probes maintain:If do not need to use the unit, please turn
it down to prevent the probe in oxidation.
3.2.9 The lifetime of hand-held pieces is 2000 hours, after the
expiration of delay, please buy replacement from my company.
3.3 touch screen
3.3.1 Do not use sharp instruments impact touch-screen.
3.3.2 Do not let cosmetic or other dirt attached to the
touch-screen
3.3.3 Do not pour liquid substance on the touch screen
3.3.4 Prohibit alcohol or detergents and other cleaning agents
scour the touch screen, For cleaning, use a soft clean cotton
cloth for light Sassafras.
4 Consumables
4.1 Conductive gel is for Consumables
4.2 The above Consumables can be bought from our company..
7. Simple malfunction and solved method
When the equipment fails during use, the user can diagnosis and
exclude some simple failures by this below guide, if still not
resolved, please call our customer service for further
consultation
5.1 Can not boot
confirmed AC is the correct access, the power plug is inserted
tight, the power switch is all open, fuse whether it is in normal.
5.2 Touch-screen invalid clicks because of below reasons:
a. May be touch screen is pressed, gently point a few on touch
screen and then reboot;
b. touch-screen positioning is incorrect or breakdown , make touch
screen calibration, please contact our customer service to solve
the problem.
5.3 feeling unwell in operation
Re-select the treatment system, ensure the applies to you.
5.4 Touch-screen appears abnormal crash
Please turn off the power supply, after 2 minutes reboot the
Machine
5.5 Abnormal situations or failure can not be excluded
Please contact the local offices, and show the purchase date, we
will be in 24 hours give reply for your question.
5.6 Replace fuse method (their position is as the figure show)
5.7.1 As shown in Figure 6.7-1 refers to the location of the arrow
type and use screwdriver to pry out the fuse holder.
5.7.2 As shown in Figure 6.7-2 put the fuse into the slot fuse
holder of installation. power supply for the AC220V, the fuse
specifications: Tac250V/8A;when use Ac110V power supply, the fuse
specification is: Tac250V/15A
5.7.3 Put the fuse holder into sockets.