Product Details

 

High Voltage Insulation Flash Tester

 

Product overview:

 

■Meet the test requirements of general electrical safety standards such as IEC 60884-1:2013 clause 17, etc
■ Up to 100 groups of test condition memory, model of each group can be set, easy to find, test data can be saved via USB(optional)
■ Voltage ramp-up and ramp-down functions to meet the requirements of new industry standards;
■ Breakdown current upper and lower limit preset, intelligent judgment;
■ Digital closed-loop feedback control mode, automatic voltage regulation mode, automatic correction of output voltage/current, and not affected by the external power grid, ensuring the accuracy of the test

■ High-definition LCD display and simple panel operation

 

Technical parameter:

 

ModelBF9636H
Test ContentsAC and DC withstand voltage, insulation resistance
Data storage methodUSB (optional)
External interfaceAlarm LAMP, PLC, RS-232
AC withstand voltage
Output voltage setting range / accuracy

AC 200~5000V

±(1.5%×setting+5digits)

Breakdown current measurement range / accuracy

0.10~100.0mA

±(1.5%×setting+5digits)

Breakdown current preset alarm range

Upper limit :0.1~100.0mA;

Lower limit :0.00~5.00mA

Short-circuited current capacity≥200mA (the device will enter into protection status when short-circuited current exceeds 200mA)
Test time range / resolution1~300s 1s/step
DC withstand voltage
Output voltage setting range / accuracyDC 200~5000V ±(1.5%× setting +5 digits )
Breakdown current measurement range / accuracy0.10~10.00mA ±(1.5%×display+5 digits)
Breakdown current preset alarm rangeUpper limit: 0.10~10.00mA; lower limit: 0.00~5.00mA
Test time range / resolution1~300s 1s/step
Insulation resistance
Output voltage setting range / accuracyDC500/1000V ±(1.5%× reading +5 digits )
Insulation resistance measurement range / accuracy

1~2000MΩ:

≤100MΩ:± (5%×display+3 digits )

>100 MΩ:± (8%× display +8 digits)

Insulation resistance alarm setting range

 

Upper limit: 0 ~ 2000MΩ; lower limit: 1 ~ 2000MΩ

Test time range / resolution

 

1~300s 1s/step

Equipment dimensions435MM(W)*150MM(H)*550MM(D)
Equipment weightApprox. 30KGS
Equipment accessoriesPower cord*1, high voltage probe*1, test wire*1, fuse*1

Hot Products

Freezing Load Test Package for IEC 62552 Refrigerator Energy Efficiency Test Overview: The ...
ISO 5151 60K BTU Computer Control Dual Code Air-Enthalpy Calorimeter System introduction: The air ...
IEC 62552 Refrigerator Freezer Energy Efficiency Lab Overview: This refrigerator performance lab is ...
ISO15502 Household Refrigerating Appliances Performance Lab 6 Stations Overview: This refrigerator ...
5RT Air Enthalpy Laboratory Air Conditioner Energy Efficiency Performance Lab System overview This ...
IEC61855 Figure 9 Dryer Air Volume Test System For Household And Similar Purposes Standards: QB...