Pressure Generator Static Pressure Range:50 mmHg-400 mmHg Resolution:1mmHg Difference between target pressure and actual pressure :±10 mmHg
from 100-400 mmHg with a minimum volume of 300 cc
Internal leak rate <2 mmHg per minute, with a minimum volume of 300 cc
Pressure Leak Test Pressure generate up to 400 mmHg (53.3 kPa) with internal pump.The
screen always show the peak pressures and the real-time
pressures.Leak rate displayed after 10 seconds.
Pressure Relief Test Pressure generate up to 400 mmHg (53.3 kPa) with internal
pump,Displays peak pressure.
User-Defined simulations Allows the operator to define blood pressure simulations. Systolic(mmHg) 20-250 Diastolic(mmHg) 10-200 Heart Rate,(bpm) 30-250 Pulse Volume,(cc) 0.1-2.4
Simulant characteristics The maximum pulse volume:2.4cc The maximum heart rate:250bpm,200bpm at pulse volume>2.00cc. Internal Adult cuff volume:20cc Internal neonate cuff volume:300cc Blood pressure units: mmHg,kPa Heart Rate Accuracy:±1BPM Patient Conditions: Weak pulse, Tachycardia,Obese,Geriatric ±1%±1BPM Mild exercise±1.5%±1BPM Strenuous exercise±3%±1BPM
Display 65k colors 4.3 inch TFT Display, brightness can be changed.
Power AC100 - 240 V 50/60 Hz, 60 VA
Environmental Conditions Working temperature: 15℃ to 40℃(59℉ to 104℉) Storage temperature: -20℃ to +65℃(-4℉ to +149℉) Relative humidity: < 90%,No condensation.