GE Datex Ohmeda Lot# 4901 Bellows Subassy Adult ABA W Disk Ring Bumpers 1500-3378-000 For Datex Ohmeda 7100 Anaesthesia

Country/Region china
City & Province guangzhou guangdong
InquireNow

Product Details

 

GE Datex-Ohmeda Lot# 4901 Bellows Subassy Adult ABA W Disk Ring Bumpers 1500-3378-000 for Datex-Ohmeda 7100 Anaesthesia machine

 

 


Brief Introduction

The Bellows assembly is the interface between drive gas and the patient circuit in the breathing system. The pressure relief valve (or pop-off valve) in the bellows assembly limits pressure in the patient circuit. Excess fresh gas is discharged through the exhalation valve into the gas scavenging system.
The Bellows Pressure Relief Valve (PRV) is normally closed, maintaining approximately 1.5 cm H2O in the breathing circuit in a no flow condition, enough to keep the bellows inflated. It is piloted closed during inspiration and remains closed until the bellows is refilled during exhalation. If the pressure in the patient circuit exceeds 4 cm H2O, the pop-off valve opens to exhaust excess fresh gas flow at a rate up to 4 L/min

 

 

 

Inspiratory ValveCalibration

1. Remove the bellows and pressure relief valve (pop-off) from the bellows assembly.

  • Remove the bellow housing.
  • Remove the bellows and bellows mounting rim as an assembly.
  • Remove the pressure relief valve.
  • Replace the bellows housing.
  • Note: If you have just completed the PEEP Valve Calibration, remove these parts from the bellows assembly and then reassemble the breathing system.

2. Attach a patient circuit tube to the Calibrated Flow Orifice test tool.

3. Connect the open end of the patient circuit tube to the inspiration flow patient connection.

4. Set the Bag/Vent switch to Vent.

5. Establish drive gas for the ventilator.

6. Adjust fresh gas flow so the actual Airway Pressure reads 1.05 cm H2

 

 

 

Breathing circuit parameters

Compliance:Bag mode:1.82 mL/cm H2O
 Mechanical mode:Automatically compensates for compression losses within the absorber and bellows assembly
Circuit volume:2.7 L Vent Mode
1.2 L Bag Mode

 

 

 

Calibration setup
1. Remove the ABS breathing system from the machine.
2. Remove the Exhalation Valve.
3. Separate the Bellows Module from the Circuit Module.
4. Install the Circuit Module only.
5. Attach a patient circuit tube to the Calibrated Flow Orifice test tool.
6. Insert the Calibrated Flow Orifice into the manifold (PEEP) port.
7. Connect a pressure sensing tee to the inspiration flow patient connection.
8. Connect the open end of the patient circuit tube to the flow port of the pressure sensing tee.
9. Connect a manometer to the pressure sensing port of the tee connector.

 

 

Quality Guarantee

Our company have one division of labour on the basis of specialization, highly cooperative group, rich technical force of medical machines development. Our expert team provides prompt, dependable solutions for sales and repairs on patient monitor, defibrillator machine, ECG/EKG machine and other accessories or repair parts. We have been always focusing on producing superior products and providing first-class customer service.

What calls for special attention is that there are both brand new and second-hand products for sell. Please remember to carefully distinguish them when placing your orders. Otherwise, it will cause unnecessary troubles. If you have any doubts about the product quality, please don’t hesitate to contact us.

 

 

Thank you for your time! If you are interested in EDAN F2 F3 F6 F9 US Transducer 8 ultrasound crystals 1 Mhz pink label cable MS3-109301(D) REF 02.01.109301 MPN 02.01.109301017 (01)06944413805484 (21)H22101060151. Please feel free to contact me, if your are interested in any of products.

Hot Products

004-0004-004 2016792-002 REV A GE Dash4000 patient monitor display assembly 10.4inch Display ...
2026653-019 GE Dash3000 Dash4000 Dash5000 Patient Monitor Battery Door Battery door features High ...
P/N 2019271-001 GE Dash3000 Dash4000 Dash5000 Patient Monitor Display Flex Flat Cable About company ...
P/N 801394-001 GE Dash3000 Dash4000 Dash5000 Patient Monitor Power Supply Board The monitor ...
P/N 2026653-037 GE Dash3000 Dash4000 Dash5000 Patient Monitor Handle The handle has a variety of ...
P/N 1076374(1058272) Philipilip V60 Ventilator Battery 14.4V 11.0Ah 163Wh LOT M91484-P1 Install an ...