ITG M-04 Medical Oxygen Sensor For Puritan Bennett™ 840 Ventilator Machine

Country/Region china
City & Province guangzhou guangdong
Categories Solar Chargers
InquireNow

Product Details

 

ITG M-04 Oxygen Sensor Medical Sensor for Puritan Bennett™ 840 Ventilator machine

 

 

Oxygen sensor replacement with access port

1.Locate the flexible oxygen sensor access cover on the top right edge of the BDU housing.

2.Firmly push the center of the lower flap of the access cover until the lower flap is dislodged from the housing.

3.Pinch the bottom and top flaps of the access cover firmly together and pull the access cover away from the housing to remove.

4. Press wire retainer tab away from the connector to release the sensor cable connector. And gently remove the connector from the oxygen sensor.

5.Remove the oxygen sensor from check valve housing.

6.Slide the new O-ring onto the threaded end of the replacement oxygen sensor. Seat the O-ring snugly against the sensor base above the threads.

7.Insert the threaded end of the oxygen sensor into the cavity on the check valve housing and finger-tighten the sensor without using excessive force. As the sensor is installed, ensure that it is not cross-threaded.

8.Connect the sensor harness connector to the oxygen sensor with the ridge on the cable connector oriented towards the wire retainer tab on the oxygen sensor.

9.Replace the access port cover by first sliding its top flap into the opening on the top of the BDU housing.

10.Using both thumbs, seat the port cover in the port by press the two outside corners of the lower flap at the juncture of the housing edge, fitting the into the housing opening.

11.Continue using both thumbs and firmly press the lower flap into place, working around the flap from the outside corners to the bottom center to seal the access cover. Be sure that the cover properly seals the housing opening.

12.Calibrate oxygen sensor by pressing 100% O2/CAL 2 min key. Verify that the calibration passes.

13.Run an SST to check the system before placing the ventilator in service.

 

 

Oxygen sensor calibration test

Test the oxygen sensor calibration as follows:

1. Connect the ventilator’s oxygen hose to a known 100% O2 source (for example, a medical-grade oxygen cylinder). Press the 100% O2/CAL 2 min key or the INCREASE O2 2 min key to calibrate the oxygen sensor. Proceed to the next step once the key light turns off.

2. Connect the ventilator oxygen hose to another known 100% O2 source (for example, a second medical-grade oxygen cylinder).

3. Set O2% to each of the following values, and allow one minute after each for the monitored value to stabilize:

21%

40%

90%

4. Watch the upper screen to ensure the value for O2 (delivered O2%) is within 3% of each setting within one minute of selecting each setting.

 

 

► Key Feature

Measurement Range:0 to 100 Vol.%
Expected Operating Life:~ 1,000,000 Vol.% h
Electrical Connector:3-pin Molex gold plated
Initial Output Signal:13.0 to 16.5 mV @ dry ambient air
Response Time t90:< 12 s
Operating Temperature:10 to 40 °C
Pressure Range:700 to 1250 hPa
Linearity Error:≤ 3 % @ 100 % O2,applied for 5 min
Zero Offset Voltage:≤ 200 μV in 100% N2, applied for 5 min
Repeatability:± 1 % Vol. O2 @ 100 % O2, applied for 5 min
Influence of Humidity:- 0.03 % rel. O2 reading per % RH
Recommended Load Resistor:≥ 10 kOhm

 

 

Thank you for your time! If you are interested in Philipilip Suresigns VM4 VM6 VM8 patient monitor Recorder Assembly Printer Module 453564015961, please let us know. Welcome to place your orders.Thank you for your time and interest.

Hot Products

Used Nonin Model 7500 Pulse Oximeter, Hospital Medical Monitoring Devices ► Brief Introduction The ...
Hospital Medical Monitoring Devices Mindray BC-2800 Auto Hematology Analyzer ► Brief Introduction ...
Hospital Medical Equipment Philipilip Glodway G30 Patient Monitor ►Brief Introduction Goldway G30 ...
P/N M3535-62301 philip MRX Defibrillator Spare Parts NBP Module NBP Module The NBP module performs ...
051-000633-00 Mindray MEC1200 MEC2000 Patient Monitor Mainboard Motherboard The main control of the ...
Mindray IMEC8 Patient Monitor Display 8.4" TM084SDHG01 M1P6563706400 The iMEC8, iMEC10 and iMEC12 ...