Philip Heartstart XL+ Defibrillator Fan Assembly
► Brief Introduction
The Philip Heartstart XL+ Defibrillator Fan Assembly is a component of the defibrillator device that helps to keep the device cool during operation. The fan assembly is designed to prevent overheating and ensure the device functions properly during emergency situations when quick and accurate defibrillation is necessary. It helps to maintain a stable temperature inside the device to ensure optimal performance and reliability.
► Removal
a While holding the Capacitor Tray with your left hand, slowly rock the Fan side to side to loosen the adhesion, being careful not to damage the sheet metal.
b When the Fan is loose, lift it off of the adhesive.
► Replacement:
1 Clean the Fans and Power Supply aluminum plates. Do not touch the surfaces once you have cleaned them.
a If you are replacing the Power Supply Assembly and reusing the old Fans, then peel the remaining adhesive off of the Fan sides.
b Wipe Fan sides with 70% isopropyl alcohol wipes and let them dry.
c Wipe the upper and lower Power Supply aluminum plates with 70% isopropyl alcohol wipes and let them dry. Make sure alcohol does not drip into the device.
2 Reinstall the lower Fan first:
a Remove one side of the adhesive backing and secure it to the side of the lower Fan,
b Position the lower Fan, so that:
– the label on the Fan’s hub is facing away from the Capacitor Tray,
– adhesive pad is facing the Power Supply, and
– the braided bundle of Fan wires is facing you and away from the Power Supply.
– the Fan is flush with the Power Supply insulator sheet.
c Peel off the adhesives backing and install the fan on the Power Supply sheet metal.Press and hold the fan in place for 10 to 20 seconds to secure it to the metal.
3 Secure the Rear Chassis in the normal Position.
4 Repeat Step 2 (sub-steps a – c) to reinstall the upper Fan
5 Install the wires into the saddle clamps:
a Install the lower Fan braided wire bundle into the lower saddle clamp.
b Install the upper Fan braided wire bundle into in three upper saddle clamps.
c Install the ECG-Out 2-wire bundle into the three lower saddle clamps. Route the ECG-Out 2-wire bundle under the Battery Connection PCA and its wires.
d Install the Network cable into all four saddle clamps.
6 Thread both Fan braided wires through the Rear Chassis grommet.
7 Connect the Fan braided wires to the Therapy PCA in any order.
8 Install the Rear Chassis Shelf.
9 Close the device.
► Specification
Item | Condition | Range |
Temperature Range | Operating | 0 to 45°C (32 to 113°F) |
Storage | 20 to 70°C (-4 to 158°F) | |
Humidity Range | Operating | 90 % Relative Humidity (RH) max. |
Storage | 90 % Relative Humidity (RH) max. | |
Altitude Range | Operating | 1014 mbar to 572 mbar (0 to 15,000ft; 0 to 4,500m) |
Storage | 1014 mbar to 572 mbar (0 to 15,000ft; 0 to 4,500m) |
If you have any questions about Philip Heartstart XL+ Defibrillator Fan Assembly , please feel free to contact us at any time, we are glad to answer them.