Philip EPIQ 5GI Ultrasound Transducer L12-3

 

 

Specification of Philip EPIQ 5GI Ultrasound Transducer L12-3

 

Product Name

Philip EPIQ 5GI Ultrasound Transducer L12-3

 

BrandPhilip
Part No

L12-3

 

Compatible withPhilip EPIQ 5GI Ultrasound Machine
TechnologyBroadband
Frequency range12 - 3 MHz
Aperture38 mm
Number of elements160
Array TypeLiner
ApplicationsVascular (carotid, arterial, and venous), intervention, bowel, MSK and small parts, and superficial imaging applications
Modes2D, steerable pulsed wave and color Doppler, color power angio (CPA), SonoCT, variable XRES, and multivariate harmonic imaging
Biopsy capableYes
Image Fusion Navigation capableNo

 

 

 

Cleaning of sensors, cables and connectors.

 

These general cleaning instructions (low-level disinfection) apply to all sensors and cables. Be sure to follow the steps below to clean the transducer and cable.

 

Caution

 

Always wear safety glasses and gloves when cleaning and disinfecting the equipment.

 

Caution

If using a premixed solution, be sure to note the expiration date of the solution.

 

Caution

 

The transmitter must be cleaned after each use. Transducer cleaning is an important step before effective disinfection or sterilization. Always follow the manufacturer's instructions when using disinfectants.

 

 

Caution

Attempting to disinfect cables or connectors using methods other than those listed here may damage the equipment and void the warranty.

 

Caution

Do not allow sharp objects (such as scalpels or welding knives) to come into contact with the transducer or cable.

 

Caution

When handling the transducer, do not strike it against a hard surface.

 

Caution

Do not use a surgical brush to clean the transducer. Even using a soft brush can damage the sensor.

 

Caution

Do not use a brush on the connector labels.

 

Caution

During cleaning, disinfection, and sterilization, place the transducer, connector, or cable components, which must remain dry, over wet components until all components are dry. This helps prevent fluid from seeping into unsealed areas of the sensor.

 

Caution

Do not allow any liquid to enter the connector. Be careful not to get liquid through the strain relief, connectors, electrical contacts, or the areas around the bolt shaft and the strain relief. Liquid in the connectors may void the device's warranty

 

Caution

You can only use alcohol solutions to disinfect the connectors. Make sure the solution contains no more than 70% alcohol. Solutions with an alcohol content higher than 70% can damage the product.

 

Caution

Do not use alcohol or alcohol-based products on cables or cable glands.

 

Caution

Avoid contact of the disinfectant with the connector labels

 

Caution

Be sure to use the correct concentration of enzymatic cleaner and rinse thoroughly

 

Caution

Do not use paper products or abrasives to clean the transducer. They can damage the soft transducer lens.

 

Caution

Make sure the transducers are completely dry before storing them. If you need to dry the transducer lens after cleaning, use a soft cloth instead of wiping.

 

 

Thank you so much for your attention .If you're interested in L12-3 Transducer Probe Philip EPIQ 5GI Ultrasound Machine ,feel free to contact us .Thank you .

Hot Products

Mindray MPM Module Pump with Solenoid Valves Assembly ► Brief Introduction The monitor uses the ...
Mindray MPM Module with IBP PN M51A-30-80873 ►Quick Detail: This product is Mindray MPM Module with ...
Mindray Nellcor OXIMAX MPM Module 115-036653-01 ►Quick Detail: Microwave Power Module (MPM), ...
Mindray MPM Module Cover Casing M51A-30-80900 ► Brief Introduction The Mindray MPM Module Cover ...
Mindray MPM Module Infrared board With IBP MPM module infrared communication back board FCF-060701 ...
Mindray MPM Module Parameter board With IBP FCF-060701 6800-20-50129 6800-30-50120 ► Brief ...