China World Technology Medical Equipment Service Group |
Verified Suppliers
|
|
This product is an adult NIBP cuff, latex-free and made of pure cotton.
What should I do if the NIBP is underinflated during blood pressure
monitoring?
When measuring blood pressure, "the cuff is too loose" or the cuff
is leaking, and the inflation pressure cannot be inflated (lower
than 150mmHg), making it impossible to measure.
Approach:
(1) There may be real air leaks, such as in the cuff, airway tube
and various joints, which can be determined through "air leakage
detection".
(2) The patient mode selection is incorrect. If an adult cuff is
used but the patient type of the monitor is a newborn, this alarm
may occur.
Troubleshooting method: Replace the blood pressure cuff with good
quality or choose a suitable type.
What should I do if the NIBP measurement value during blood
pressure monitoring is inaccurate?
The measured blood pressure value deviates too much.
Treatment method: Check whether the blood pressure cuff is leaking,
and whether the pipe interface connected to the blood pressure is
leaking. Or is it caused by the difference in subjective judgment
between the auscultation method and the auscultation method?
Troubleshooting method: Use the NIBP calibration function. This is
the only standard available to verify that the NIBP module
calibration values are correct at the user's site. The standard
deviation of the pressure tested by NIBP before leaving the factory
is within 8mmHg. If it exceeds, the blood pressure module needs to
be replaced.
1. Adopt one-piece bagless design, smooth and comfortable
2. Adopt humanized rounded corner design, beautiful appearance and
comfortable to use.
3. With different types of connectors, the cuff can be applied to
various monitor series
4. For adults to meet patient needs
5. Double tube design
6. Latex-free and biocompatibility tested
7. Comply with ANSI/AAMI SP10-2002, EN60601-1:1990
Routinely measure upper arm blood pressure. Routine measurement of
wrist blood pressure and finger blood pressure is not recommended.
① It is recommended to measure the blood pressure of the left and
right upper arms (brachial artery) for the first time, and use the
side with higher blood pressure as the upper limb for blood
pressure measurement.
②When the difference in blood pressure (systolic blood pressure)
between the left and right upper arms is >20 mmHg, it is
recommended to measure blood pressure in the limbs.
③The elderly and patients with diabetes or certain diseases are
prone to orthostatic hypotension, so measure blood pressure in
multiple positions. Blood pressure can be measured in the supine or
standing position when necessary. The standing blood pressure
measurement should be performed 3 minutes after the patient changes
from lying to standing.
Model | Adult NIBP Cuff Double Hose |
certification | CE |
LATEX | NO |
Material | cotton |
Size | 25-35cm |
Warranty | 90 days |
Condition | Good |