Philip MX800 Patient Monitor LCD Display LQ190N1LW01 Original New

Country/Region china
City & Province guangzhou guangdong
InquireNow

Product Details

Philip MX800 patient monitor LCD display LQ190N1LW01 original new

 

 

About Philip MX800 patient monitor LCD display LQ190N1LW01 original new

 

The Philip MX800 patient monitor features a high-resolution LCD display that provides clear and detailed monitoring data for healthcare professionals. The display is bright and easy to read, even in low-light conditions, making it easy to monitor patient vital signs and make informed clinical decisions. The MX800 monitor also allows for customizable display options, so healthcare providers can tailor the information presented on the screen to meet their specific monitoring needs. Overall, the LCD display on the Philip MX800 patient monitor helps healthcare professionals to effectively monitor and care for their patients.

 

 

Philip MX800 patient monitor LCD display LQ190N1LW01 original new Specification

 

Product Name

Philip MX800 patient monitor LCD display LQ190N1LW01 original new

 

BrandPhilip
Part noLQ190N1LW01
ApplicationPhilip MX800 patient monitor
Size19”
Type482 mm active matrix color LCD (TFT)
Resolution1680 x 1050 (WSXGA+)
Refresh frequency58 Hz
Useful screen409.5 mm x 255.9 mm
Pixel size0.244 mm x 0.244 mm

 

 

Troubleshooting

 

Integrated Display is blank

 

SymptomsPossible Causes of FailureFailure Isolation and Remedy
Integrated display is blank or brightness is reduced (The information listed in this table is only valid if the boot phase has completed without error. See table for a description of the Boot phase.)Display brightness is reduced when room temperature, or instruments placed near patient monitor, causes the monitor display to overheat.Instrument should be placed in an environment that does not exceed 40 degrees C or below 5 degrees C.
If you have an external display, connect it to the video port. If the external display works, you can eliminate the connector board and the main board as the cause of failure.

 

 

 

About Philip MX800 patient monitor

 

The IntelliVue MX800 Patient Monitor:

 

• View real-time data.

• Control connected measurement servers.

• Notify if there is a problem with patient or equipment.

• Provides limited storage and retrieval space (trend).

• Connect to the Philip clinical network and other devices. Monitors with a single integrated measurement server can be connected to other basic components. To create a monitoring system with a large number of measurements. Additional interface capabilities and multiple slave displays These components work together as an integrated real-time measurement system.

 

 

 

Thank you so much for your attention .If you're interested in Philip MX800 Patient Monitor LCD Display LQ190N1LW01 Original New ,feel free to contact us .Thank you .

Hot Products

Mindray Pulse Oximeter PM-60 Brief Introduction of Mindray Pulse Oximeter PM-60 The PM-60 is a ...
Sino-K Medical SPH100 Handheld Spo2 Finger Tip Pulse Oximeter product features: The working ...
Olympus EVIS LUCERA CV-260 Endoscopy Endoscopy Function 1. Suction connector Connect the endoscope ...
Biolight BLT AnyView A5 Patient Monitor Modular Buckle ► About Biolight BLT AnyView A5 Patient ...
Mindray BIS LOC 2 Channel Cable Assembly 6800-30-50761 185-1014-MR 0010-21-42759 ► Brief Introductio...
Mindray BeneHeart D6 Defibrillator Machine ► Brief Introduction The Mindray BeneHeart D6 Defibrillat...