Drager Neonatal Straight Flow Sensor ISO15 8411130 for Straight flow sensor for Babylog® 8000/Evita® 4/ Evita 2 dura®/Evita® NeoFlow.

 

 

 

►Brief Introduction

The measurement principle used in flow measurement is the hot wire velocimetry method.Hot wire velocimetry is a thermal measurement method in which the measuring line of the flow meter remains at a constant high temperature during the measurement process. The higher the flow rate, the more energy is required to maintain a constant high temperature. The flow rate can then be calculated from the energy consumed. To ensure proper functionality, check for visible damage, dirt and particles before inserting the flow meter. Repeat this check regularly. If the flow meter is damaged, dirty, or contains particles, replace it. If the measurement line of the flow meter continues to glow during operation, it is a sign of dirt. Reprocess or replace the flow meter immediately.

 

Work for Draeger Inc.

  • BABYLOG 8000
  • EVITA 4
  • EVITA XL
  • BABYLOG
  • EVITA 4/2D

 

 

► Safety-related information

Flow sensors

Flammable substances

The flow sensor may ignite medications or other substances based on easily flammable substances. The patient may be put at risk.

  • Do not nebulize medications or other substances that are easily flammable or spray them into thedevice.
  • Do not use substances containing alcohol.
  • Do not allow combustible or explosive substances to enter the breathing systemor the breathingcircuit.

 

Residual vapors of highly flammable disinfectants

Residual vapors of highly flammable disinfectants (e.g., alcohols) and deposits that were not removed during reprocessing may ignite when the flow sensor is in use.

The patient may be put at risk.

  • Ensure particle-free cleaning and disinfection.
  • After disinfection, allow the flow sensor to air-dry for at least 30 minutes.
  • Before inserting, check the flow sensor for visible damage and soiling, such as residual mucus,medication aerosols, and particles.
  • Replace flow sensors when damaged, soiled, or not particle-free.Neonatal flow sensor improper reprocessing and soiling, such as deposits or particles, may damage the flow sensor. The flow measurement may fail. As a result, the patient may be put at risk.
  • No machine cleaning or disinfection of the sensor insert
  • No plasma sterilization or radiation sterilization
  • No water jets, compressed air, brushes or the like when cleaning the sensor insert
  • No ultrasonic bath

 

 

► Steps

1. Pull out the flow meter (6) from the Y-piece (5).

2. Unplug the flow meter cable plug (7) from the flow meter (6).

3. While gently pressing the knobs (8) on both sides, pull the flow meter insert (9) out of the flow meter housing (10).

 

Specification

ItemNEONATAL FLOW SENSOR ISO15
OEMDraeger Inc.
OEM #8411130
ReturnableNo
StandardsISO 15

 

 

 

If you have any questions about Drager Neonatal Straight Flow Sensor ISO15 8411130 for Straight flow sensor for Babylog® 8000/Evita® 4/ Evita 2 dura®/Evita® NeoFlow., please feel free to contact us at any time, we are glad to answer them. Looking forward to receiving your early inquiry. Thanks for Your Cooperation.

 

Hot Products

Philip Goldway UT6000A Patient Monitor Main Board Mother Board M-6A0S01B ► Brief Introduction The ...
Philip Efficia CM150 Patient Bedside Monitor Used For Hospital Proven physiological measurement ...
Comen K12 Patient Monitor Comen K12 Patient Monitor 12.1 inch Medical Equipment For Hospital /Clinic ...
Edan IM8 patient monitor keyboard MS1R-16989-V1 ► About Edan IM8 patient monitor keyboard MS1R-16989...
Comen V1 Ventilator Machine Comen Portable And Compact V1 Ventilator Machine Used For Hospital About ...
Comen K1 patient monitor Hospital Medical Equipment Comen K1 Patient Monitor In Good Working ...