Mindray D3 Defibrillator Paddle Plate Electrode Pads

 

 

 

SymptomPossible CauseCorrective Action

 

 

 

The equipment cannot be charged by pressing the Charge button on the external paddles

Paddles not connected properlyReconnect the paddles
Paddles failure.1. Isolate the problem by connecting pads to perform charging/discharging. 2. If the paddles are failed, replace them
Failure of therapy module.Replace the therapy module
Failure of connection wire to the therapy module.Replace the connection wire.
A shock cannot be delivered by pressing the Shock button on the paddles.Paddles not connected properly.Reconnect the paddles
Paddles failure.

1. Locate the problem by connecting pads to perform charging/discharging. If normal discharge can be performed, it indicates the paddles are defective.

2. Replace the paddles.

Failure of therapy module.Replace the therapy module
Connection wire to the therapy module broken.Replace the connection wire.
The equipment can be properly charged, but the energy is disarmed automatically at the completion of charging or when the equipment is being discharged.

Too high or too low patient impedance detected.

 

1. Pads/paddles are detached from the patient.

2. Pads/paddles failure.

3. Pads cable failure

Ensure good connection between the patient and pads/paddles. If the problem persists, replace Pads, paddles or Pads cable.
Failure of therapy module.Replace the therapy module.
Defibrillation hardware circuit defectiveReplace the therapy module.
Energy Select buttons on the paddles do not workPaddles not connected properly.Reconnect the paddles.
Paddles failure.Replace the paddles.
Failure of therapy module.Replace the therapy module.
Connection wires to the therapy module brokenReplace the connection wires.

 

 

 

Company Introduction

 

CWT Technology HK Limited is a second-hand medical equipment manufacturer based in Guangzhou. Our main products are used patient monitors, electrocardiograms, defibrillators, ultrasound or their parts and accessories. We also have breathing apparatus, spare parts and accessories for anesthesia machines. And always strive to improve customer satisfaction. Meeting or exceeding the standards and requirements of our customers and manufacturers We never stop educating ourselves about business ethics and professional standards. Our engineering, sales and management teams provide professional customer service and quality technical support around the clock. Seven.

 

The company's service principle is customer first, quality assurance. Honesty!

 

Company business goals: Our business is based on integrity. We provide our customers with the latest medical equipment. Contributing to the development of the medical industry

 

 

Thank you so much for your attention .If you're interested  in Mindray D3 Defibrillator Paddle Plate Electrode Pads,feel free to contact us .Thank you .

Hot Products

Philip Goldway G30 patient monitor alarm light board C-ALARM012A JM292470503 ►Brief Introduction ...
Philip Goldway G30 Patient Monitor Battery Board M-G301J02B In Good Working Condition Philip Goldway ...
Philip Goldway G30 Patient Monitor Battery Door Assembly Hospital Medical Equipment In Good Shape ...
Philip Goldway G30 patient monitor display and connector cable ►Brief Introduction Those cables is ...
Philip Goldway G30 patient monitor system board M-4B3S01A Hospital Medical Equipment Philip Goldway ...
Hospital Equipment Machine Biolight BLT AnyView A5 Patient Monitor Bedside Monitor In Good Working ...