Touchscreen Calibration

 

To access the touchscreen calibration screen:

 

1 Enter service mode

 

2 Select Main Setup

 

3 Select Hardware

 

4 Select Touch Calibration

 

 

Perform the calibration procedure without turning off the monitor. If the screen turns off after touching the first point, the touch screen will be disabled until touch calibration is performed again.

If you calibrate the touchscreen incorrectly by selecting the wrong location, you will need to re-enter calibration mode using a different input device. If you have a support device, you can choose to reset touch calibration to the default. This creates an approximate calibration that allows you to return to the calibration menu from the touch screen.

 

 

Integrated Touch Display not functioning

 

1.Touch Screen not functioning

Possible Causes of FailureFailure Isolation and Remedy
The touchscreen function is temporarily disabled

Check if the touch screen function is temporarily disabled (lock symbol on the home button). If yes, press and hold the main button

The on-screen button reactivates the touchscreen.

The touchscreen function is permanently disabledIn Service Mode, select Main Settings -> User Interface and change "Enable Touch" to "Yes".
Tablet adapter not connectedCheck the connection between the tablet adapter and the motherboard
The touch controller is not connectedCheck the connection between the touch controller and the tablet adapter
Touch screen cable not connectedCheck the connection between touch control panel and touch panel
Defective touch control board or touch sensor

Touch control panel and touch sensor replacement

Note: Linearization data must be uploaded and recalibrated later

Touch control panel replacement

the motherboard is brokenReplace motherboard

 

2.Invalid touch position

 

Possible Causes of FailureFailure Isolation and Remedy
the touch is not calibrated

To calibrate by touch:

1. Enter the main settings menu

2. Select hardware

3. Select the touch adapter settings

4. Select Calibrate

 

Disable/enable touch actions

There are two ways to disable/enable the touchscreen functionality:

1. Press and hold the home button to temporarily disable the touchscreen functionality. A padlock symbol appears on the key. Press and hold the home button again to reactivate the touchscreen function.

2. To permanently disable the touchscreen:

A. Enter service mode.

B. Select main settings

C. Select User Interface

D. Change Enable Touch to no.

Set Enable Touchscreen to Yes to re-enable touchscreen functionality

 

Removing the Touchscreen

 

1 Remove the front and back of the monitor as above.

2 Remove the touch controller board, press the connector safety latch, and disconnect the touch sensor cable.

Note Both the touch screen and touch controller board must be replaced for the monitor to function properly.

Assembly Note: First connect the touch sensor cable to the touch controller board, then reinstall the touchpad.

 

3. Remove the two screws securing the power button and LED board and remove the board. Plates and screws are part of the flex panel adapter cable.

Note The screws that secure this plate are self-tapping screws. When reinstalling, first turn it in the opposite direction.

4. Remove the 8 screws, loosen the 2 clips on the monitor frame and carefully remove the monitor assembly.

NOTE: It is recommended that this procedure be performed in a dust-free environment. The touchscreen display assembly can be difficult to remove due to the gaskets (tape) holding the two together.

5 Remove the touch screen (sensor) from the frame.

Note This operation can be difficult due to the seams (bands).

Assembly Note: When viewing the touch screen from the rear, the touch sensor cable must be facing to the right when placing the touch screen back into the monitor housing.

Note Both the touch screen and touch controller board must be replaced for the monitor to function properly.

 

If you're interested in philip MP40 Patient Monitor Refurbished Touch Screen E131992 J09L078118 ,feel free to contact us .Thank you .

Hot Products

philip VM6 LVDS cable 10 philip VM6 LVDS cable Medical Equipment For Hospital Product Description: ...
GE Dash 4000 Patient Monitor GE Healthcare DASH 4000 Patient Monitor Medical Equipment Good Working ...
philip CAPNOSTAT M2501A Patient Monitor CO2 Sensor philip CAPNOSTAT M2501A Patient Monitor CO2 ...
COMEN C60 Neonatal Patient Monitor Used for Hospital About COMEN C60 patient monitor COMEN C60 has ...
Used Mindary VS-600 VS600 Vital Signs Patient Monitor for Adult Pediatric Neonatal ❖ Mindray VS-600 ...
Medical Fetal Monitor Parts philip FM20 Fetal Monitor Cover Protector M2703-44150 philip FM20 Fetal ...