Model | DZS-706 | DZS-706-A | DZS-706-B | DZS-706-C |
Parameters | Cond/DO/pH/pX/℃ | Cond/ pH/pX/℃ | DO/pH/pX/℃ | Cond/DO/℃ |
Range | pH/ pX | -2.00~20.00 | — |
ORP | ±1999.9mV | — |
Conductivity | 0.000μS/cm~199.9 m S/cm | — | 0.000μS/cm~199.9 mS/cm |
Resistivity | 0.00Ω·cm~20.00MΩ·cm | — | |
TDS | 0.000 mg/L~100g/L | — | 0.000 mg/ L~100g/L |
Salinity | 0.00~8.00% | — | 0.00~8.00% |
DO | 0.00~19.99mg/L | — | 0.00~19.99mg/L |
DO Saturation | 0.0~199.9% | — | 0.0~199.9% |
Response Time | ≤45s (90% respond at 20℃ ) | — | ≤45s (90% respond at 20℃ ) |
Ion Concentration | (0~19990), unit: ug/L,mg/L,g/L,mol/L,mmol/L | — |
Temperature | -5.0~105.0℃ |
Resolution | pH/ pX | 0.01 | — |
ORP | 0.1mV | — |
Conductivity | 0.001μS/cm | — | 0.001μS/cm |
Resistivity | 0.01Ω·cm ; | — | 0.01Ω·cm |
TDS | 0.001mg/L | — | 0.001mg/L |
Salinity | 0.01% | — | 0.01% |
DO | 0.01mg/L | — | 0.01mg/L |
DO Saturation | 0.10% | — | 0.10% |
Temperature | 0.1℃ |
Accuracy | pH/ pX | ±0.01pH±1 bit | — |
ORP | ±0.1%FS | — |
Conductivity | ±1.0%FS | — | ±1.0%FS |
Resistivity | ±1.0%FS | — | ±1.0%FS |
TDS | ±1.0%FS | — | ±1.0%FS |
Salinity | ±0.2% | — | ±0.2% |
DO | ±0.30mg/L | — | ±0.30mg/L |
DO Saturation | ±10.0%FS | — | ±10.0%FS |
Temperature | ±0.3℃ ±1 bit |
Temperature Compensation | Auto, pH:(0.0~100.0)℃ ; Conductivity: (0.0~100.0)℃ ; DO:
(0.0~40.0)℃ |
Zero Offset & Full Scale Calibration | Yes | — | Yes |
Data Storage | 2200 sets of data |
View, delete and print | Yes |
Display | LCD |
Communication Interface | USB |
Software | 1.1DC |
Power | Universal power adapter (9V DC, 800mA) |
Dimension & Weight | 280×215×93mm, 1kg |
Standard Configuration | DJS-1C Platinum Black Conductivity Electrode; E-201-C pH
Combination Electrode; DO-958-S DO Probe; Standard Buffer |