Home Companies SKYLINE INSTRUMENTS CO.,LTD

ASTM D92 Cleveland Open Cup Flash Point Tester

SKYLINE INSTRUMENTS CO.,LTD

Contact Us

[China] country

Trade Verify

Address: No.1802 Room 6# Building Universal Commerce Center Nancheng District Dongguan Guangdong China

Contact name:Ivy Xie

Inquir Now

SKYLINE INSTRUMENTS CO.,LTD

Verified Suppliers
  • Trust
    Seal
  • Verified
    Supplier
  • Credit
    Check
  • Capability
    Assessment

ASTM D92 Cleveland Open Cup Flash Point Tester

Country/Region china
City & Province dongguan guangdong
Categories Other Metals & Metal Products
InquireNow

Product Details

ASTM D92 Flash Point Oil Analysis Equipment Open Cup Flash Point Tester

 

Description Of Opened Cup Flash Point Tester                    

The fully automatic open flash point tester is designed and manufactured in accordance with the requirements specified in the People's Republic of China standard GB/T3536-2008 Determination of Flash Point and Ignition Point of Petroleum Products Cleveland Open Cup Method. It is applicable to the determination of the flash point and ignition point of petroleum products other than fuel oil, whose open cup flash point is higher than 79 ℃ according to the method specified in this standard.

 

This instrument is a fully automatic open flash point tester, which is an ideal substitute for similar imported instruments. It can be widely used in railway, aviation, electric power, oil industry and scientific research departments.

Features                           

1. It adopts 8-inch true color LCD, touch screen and Chinese man-machine dialogue interface. It has prompt menu and guided input mode for expected flash point, sample label, atmospheric pressure, test date and other parameters.

2. The function curve of temperature rise and test time is simulated and tracked, with Chinese misoperation software prompt modification function, test date, test time and other parameters prompt function.

3. Equipped with a standard 232 interface or USB interface to connect with the computer to automatically upload data. It can be connected to the LIMIS system.

4. 120 groups of data can be stored, which can be connected with the computer to store data in large capacity and for a long time.

5. Automatically correct the influence of atmospheric pressure on the test and calculate the correction value.

6. Differential detection, automatic correction of system deviation.

7. The data scanning, detection and printing are completed automatically, and the test arm is raised and lowered automatically.

8. Automatic air cooling

 

Specifications                 

Working power supplyAC (220 ± 10%) V, 50Hz
Temperature measurement rangeroom temperature~400 ℃;Temperature display unit0.1 ℃
Repeatability≤ 8 ℃ (flash point and fire point)
Reproducibility≤ 17 ℃ (flash point), ≤ 14 ℃ (fire point)
Heating ratein accordance with GB/T3536-2008 standard
Ignition modegas flame, electronic ignition, optional
Ambient temperature(10~40) ℃
Relative humidity≤ 80%
Power consumption of the whole machineno more than 500W
Boundary dimension(when the lifting arm is not lifted) 520mm × 360㎜ × 310 ㎜ (long × wide × High); (When the lifting arm is lifted) 520mm × 360㎜ × 420 ㎜ (long × wide × High)

 

 

 

 

 

 

 

 

 

 

Hot Products

SAE J2302 Test Apparatus For Thermal Effectiveness Of Sleeve Insulation Product introduction This ...
ISO 340 Conveyor Belt Vertical Burning Test Chamber 1. Scope: This tester is suitable for conveyor ...
Building Material Flammability Test Furnace with Touch Screen Control Description The building ...
Building Material Flammability Test Furnace with Touch Screen Control Description The building ...
ISO5660 Cone Calorimeter Analysis Instrument For Heat Release Product introduction Cone Calorimeter ...
ISO5660 Cone Calorimeter Analysis Instrument For Heat Release Product introduction Cone Calorimeter ...