Energy Dispersive X-ray Fluorescence Spectrometer

 

Introduction

ICP 2060T Sequential Inductively Coupled Plasma Emission Spectrometer is designed to measure major, minor and trace elements in various samples with outstanding performance.
It was widely chosen in different industries for its superior optical resolution, high level of automation, reliable free running RF system as well as excellent analytical precision and accuracy.
ICP2060T was widely applied in various fields ranging from rare earth, geology, metallurgy, chemistry, environmental protection, clinical medicine, petroleum products, semiconductors, foods, biological samples, criminal science, to agricultural research, etc.

 

Applicated fields

Rare earth metal
Silicon materials, magnetic materials processing industry
Metallurgical industry: analysis of impurity elements influencing the quality of metal materials
Water quality control:
Geology, Mineral
Petrochemical: measurements of more than 30 elements in crude oil
Pharmacy, hygiene, agriculture, environmental protection, commodities and food industry

 

 

Features

  • Over 70 elements can be measured by ICP 2060T Machine
  • Fast analysis, 5-8 elements per minute;
  • Excellent detection limits, at ppb level for most elements
  • Wide linear dynamic range, reaching 6 orders of magnitude, from ppb to percentage;
  • Lower gas consumption; Each argon gas cylinder can be used for 8 hours;
  • Advanced optical design with a fully illuminated holographic grating, with spectral interference correction and high light throughput for improved detection limits
  • Wide wavelength range: 190nm to 500nm with 3600 line grating;190nm to 800nm with 2400 line grating;
  • Computer controlled plasma platform optimizes the viewing position to reduce interferences, improve SNR and minimize
  • background emissions
  • Software controlled flow of carrier gas, plasma gas and auxiliary gas provide real time monitor as well as highly stability of flux.
  • Various types of plasma torches , nebulizer as well as spray chamber are optional.
  • Choice of auto sampler greatly improves efficiency
  • Sample waste drained by peristaltic pump ensures stable sample flow rate; Fast pump mode improves productivity
  • The brand new robust free-running, 27.12 MHz RF generator that delivers unsurpassed performance
  • Rapid and accurate automatic coupling system ensures the ultimate power transfer efficiency and stability
  • One-button plasma ignition
  • By simply click the button ‘inflame’ in software , the plasma can be ignited immediately
  • The ‘graph diagnosis’ function provides information on instrument status and analysis progress (as figure 3)
  • Multi-element analysis:after selecting elements and setting measuring parameters, instrument will measure automatically with results directly displayed. (as figure 4)
  • Enhanced database management:the database contains several thousands of spectral line, where users can choose freely based on application method. (as figure 5)
  • Strong and User-friendly Software:the professional software provide excellent features and multi access to functions as follow:

          Data management, quantitative and qualitative analysis, test parameter setting, one button report generation, background and interference correction, fast calibration mode, instrument status indication, and online self diagnostic.

 

 

Technical Specifications

1Size1.5m×0.8m×0.8m
2Frequency27.12MHz
3Frequency stability<0.05%
4Couple methodAuto match
6Output power800W ~1600W, continuously adjustable with power efficiency more than 65%
7Output power stability≤0.05%
8Induction coilID 25mm×3, equipped with three concentric quartz torch tubes of ED 35mm (ID: internal diameter; ED: external diameter)
9Spray chamberScott double pass spray chamber
10Optical typeCzery turner
11Focal length1000mm
12Grating specificationhuge holographic grating with 3600L/mil or 2400 L/mil and 80 mm×100mm of ruling area
13Resolution≤0.015nm (3600 line grating), ≤0.030nm (2400 line grating)
14Wavelength range195~500 nm for 3600 line grating, 195~800 nm for 2400 line grating.
15Photomultiplier tube specificationR212/R928
16Negative high voltage(-50 ~ -1000) V
17Circuit measuring range(10-12 ~ 10-4) A
18Signal acquisitionVF conversion


 

Computer

Lenovo PC
Monitor:17 Inch LCD
Printer:Canon Inkjet
Main power supply:AC 220V, Current: 20A

 

 

Elements Analysis 

1) Suitable sample content range :
a. Liquid sample:0.01ppm~ several thousands ppm
b. Solid or powder sample:0.001% ~ 70%
2) Repeatability:short-term stability, RSD≤1.5%
3) Long-term stability:RSD≤2%
4) Test speed:5~8 elements/min
5) Limits of detection (LOD, μg/L) for typical elements:1ppb~10ppb

 

 

Other accessories

Windpipe
Voltage stabilizer
Gas pressure regulator
ICP 2060T power cable
Water cooling system
Copper cable for ground
Sampling system include nebulizer, spray chamber, plasma torch

 

Working environment

Temperature for storage and transportation:15℃-25℃
Humidity for storage and transportation:≤70%
Power stability:220±10v 50-60Hz

Hot Products

ISO 8142 Thermal Insulation Materials Maximum Use Temperature Test Apparatus Introduction: This ...
Lateral Flame Spread Testing For Walls Surfaces Introduction Lateral Flame Spread Testing Equipment ...
Fire Propagation Index Testing Equipment for Laminated Foil BS476-6 FPT.pdf Introduction: BS 476-6 ...
ISO9239 FRPT.pdf Introduction: Critical Radiant Flux is the main parameters of the fire risk for ...
Introduction This Fire Test Equipment is developed base on the standard AS1530.2, test for ...
Introduction: The BS 476-12 Ignitability Tester is designed according to the BS476-12:1991. Method ...