Beijing Samyon Instruments Co., Ltd. |
|
Summary
This instrument is designed and manufactured as per standards GB/T
261-2008 Determination of Flash Point - Pensky-Martens closed cup
method and ASTM D93 Standard Test Methods for Flash Point by
Pensky-Martens Closed Cup Tester. It is used to make determination
of the closed cup flash point of the petroleum products.
I. Main technical features
1. Simulation tracking displays function curves of heating up and
test time.
2. Equipped with the English prompt function of mis-operation.
3. Equipped with prompt function of test date,test time and other
parameters.
4. Automatically calibrate the effect of atmospheric pressure to
test and calculate the corrected value.
5. Differential coefficient detection.Automatically correct the
system deviation.
6. Automatically open the lid,detect the flash point and print test
data.
7. The test arm rises up and lowers down automatically.
8. Igniting with electric igniting gun,forced air cooling.
II. Main technical specifications
Power supply | AC( 220±10% )V, 50Hz |
Flash point determination | ange: Ambient to 300℃. Repeatability: ≤ 2℃ Reproducibility: ≤ 4℃ Accuracy: 0.1℃ |
Heating rate | Procedure A: (5~6)℃/min |
Stirring rate | Procedure A: (90~120)RPM, Procedure B: (250±10)RPM Automatic control and manually adjustable. |
Igniting mode | (1) Igniting source: gas(or other civilian fuels,the same below) |
Ambient temperature | ≤ 40℃ |
Relative humidity | ≤ 80% |
Maximum power consumption | 500W |