Home Companies Dongguan Quality Control Technology Co., Ltd.

ASTM D256 Plastics Izod and Charpy Pendulum Impact Tester with LCD for Non Metal

Dongguan Quality Control Technology Co., Ltd.
Active Member

Contact Us

[China] country

Address: Room 301, Building No.1, No. 2 Jiaolian Yiheng Rd, Wanjiang District, Dongguan, China, 523046

Contact name:Raymond Chung

Inquir Now

Dongguan Quality Control Technology Co., Ltd.

ASTM D256 Plastics Izod and Charpy Pendulum Impact Tester with LCD for Non Metal

Country/Region china
City & Province dongguan guangdong
Categories Laser Pointers
InquireNow

Product Details

 

ASTM D256 Plastics Izod and Charpy Pendulum Impact Tester with LCD for Non Metal

 

Model: XJC-5, XJC-5D

Overview:

iqualitrol XJC Pendulum Impact Tester is used to test the anti-impact intensity both in izod and charpy of engineering plastics, glass-steel, ceramics, casting stone, plastic wiring, insulating material, plastic pipe material and other non –metallic materials.

 

iqualitrol XJC Pendulum Impact Tester equips with electronic controlling system which is much easier operated and the test result is calculated automatically.

 

Izod impact testing is an ASTM standard method of determining the impact resistance of materials. A pivoting arm is raised to a specific height (constant potential energy) and then released. The arm swings down hitting a notchedsample, breaking the specimen.

 

The energy absorbed by the sample is calculated from the height the arm swings to after hitting the sample. A notched sample is generally used to determine impact energy and notch sensitivity.

 

The Charpy impact test, also known as the Charpy V-notch test, is a standardized high strain-rate test which determines the amount of energy absorbed by a material during fracture. This absorbed energy is a measure of a given material's notch toughness and acts as a tool to study temperature-dependent ductile-brittle transition. It is widely applied in industry, since it is easy to prepare and conduct and results can be obtained quickly and cheaply. A disadvantage is that some results are only comparative.

 

Standards:

1, DIN 53453 Testing of Plastics Impact Flexural Test, The purpose of the impact flexural test is to ascertain the behaviour of specified test specimens under defined conditions of impact flexural loading, and to assess the toughness or brittleness within the limits set by the test conditions.

2, ASTM D256 - 10 Standard Test Methods for Determining the Izod Pendulum Impact Resistance of Plastics

3, ISO 179-1:2010 Plastics -- Determination of Charpy impact properties -- Part 1: Non-instrumented impact test

4, ISO 180:2000 Plastics -- Determination of Izod impact strength

 

Specification:

Velocity at Impact (m/s)Charpy: 2.9 3.8;
Izod: 3.5
Impact Energy (J)Charpy (0.5), 1, 2, 4, 5, 7.5, 15, 25, (50)
Izod (0.6), 1, 2.75, 5.5, 11, 22
Angle160°
Charpy Tip RadiusR=2mm ±0.5mm
Izod Tester Blade RadiusR = 0.8 + - 0.2 mm
Jaw RadiusR=1mm ±0.1mm
Blade Angle30°±1°
Charpy Support Spacing40mm, 60mm, 70mm, 95mm
Izod Support Spacing22mm±0.5mm
DisplayMechanical or LCD
Effective Length of Pendulum380 mm, 322 mm
Power SupplyAC220V 50Hz
 

 

Test Piece ShapeLength CLength bLength dCharpy Support SpacingIzod Support Spacing
150±16±0.24±0.24022
280±210±0.54±0.260
3120±215±0.510±0.570
4125±213±0.513±0.595
 

Models:

ModelEnergy (J)Velocity (m/s)DisplaySize (mm)Weight (kg)
XJC-5

Charpy: 1, 2, 4, 5

Izod: 1, 2.75

2.9

3.5

Mechanical500×350×800120
XJC-5D

Charpy: 1, 2, 4, 5

Izod: 1, 2.75

2.9

3.5

LCD500×350×800120
 

0.5J, 0.6J, 50J must be customized.

 

Hot Products

Cold Balanced Control System Programmable Temperature and Humidity Environmental Test Chamber ...
Constant Temperature Humidity Incubator for Pharmaceutical and Chemical Industry Product introductio...
Programmable Temperature Alternative Test Chamber QTL-330C by Forced Air Cooling Rate 5℃/min Model: ...
Digital Smart Controller Large Industrial Trolly Drying Oven for Electroplating Industry Main ...
Constant Temperature Medical Thermostatic Incubator with Perfect Air Current Cycling Thermostatic ...
Double Layer Isolation Programmable PID Control Biochemical Incubator for Drug Testing Biochemical ...