Weight (not including probe):173g
Standard accessories:
Main unit
Zero Calibration plate
Carrying case
Instruction Manual
Certificate
CALIBRATION CHECK
To check whether the tester is accurate, just Insert the indenter
into the hole of the calibrated test Block. Apply enough force to
make firm contact between the top surface of the test block and the
presser foot. The reading
should agree with the value stamped on the test block. If not, just
carry out Zero calibration and High end calibration.
Zero calibration
Hold the durometer vertically with the point of the indenter
hanging in the air, the reading on the display should be '0'. If
not, depress the 'ZERO' key to make the tester display '0' .
High end calibration
Just place the indenter onto a flat glass, apply enough force to
make firm contact between the glass and the presser foot. The
readings on the display should lie between 99.5 and 101. If not,
press CAL key to carry out high end calibration.
BATTERY REPLACEMENT
When the battery symbol appears on the display, it is time to
replace the batteries.
Slide the Battery Cover away from the tester and remove the
batteries.
Install batteries paying careful attention to polarity.
NOTE
Readings below 10HD for Shore D type may be inexact and
should not be reported for some materials. Measurements should be
made on a Shore A type.
Readings above 90HA for Shore A should be made on a Shore D type
durometer