1.Principle

KFDO310 integrated on-line fluorescence dissolved oxygen sensor is designed and made based on the quenching principle of excited fluorescence of specific substances in physics. The blue light from the light-emitting diode illuminates the fluorescent material on the inner surface of the fluorescent cap. The fluorescent material on the inner surface is excited and emits red light. By detecting the phase difference between the red light and the blue light, and comparing it with the internal calibration value, the concentration of oxygen molecule can be calculated, and the final value can be output by temperature automatic compensation.

2.Feature

No electrolytes, no polarization

No need to consume oxygen, not affected by the flow rate

Built-in temperature sensor, automatic temperature compensation

Free from chemicals like sulfides

Small drift, fast response, more accurate measurement

Maintenance-free, long service life, lower use cost

Fluorescent caps are easy to replace

RS-485 interface, Modbus-RTU protocol

3.Technical Specifications

Model numberKFDO310
Measuring principlefluorescence
Range0ー20 mg/L (0ー200% saturation, 25 °C)
Resolution0.01 mg/l, 0.1 °C
Precision± 2% f.s. , ± 0.5 °C
Temperature compensationAutomatic temperature compensation (PT1000)
Output modeRS-485 bus, Modbus-RTU protocol
Working conditions0ー45 °C, < 0.2 mpa
Storage temperature- 5 ~ 65 °C
Installation modeImmersion mounting
Cable length5 meters, other length can be customized
Power consumption< 0.05 W
Power supply12 ~ 24 VDC ± 10%
Protection levelIP68
CalibrationTwo-point calibration
Fluorescent cap lifeGuaranteed Use for one year (under normal use)
Material for sensor housingPom and 316L stainless steel
 

4.Dimensional Drawing

 

 

 

 

 

5.Installation and Electrical Connection

Installs

The temperature sensing part should be immersed below the liquid surface to avoid collision with the film head surface. The head part of the membrane should be free from sediment.

 

 

Hot Products

1. Description KPS110 pressure and temperature integrated transmitter is a temperature and pressure ...
Kacise IP66 High Accuracy Wide Pressure Range Digital Pressure Gauge With Data Logger 1. Features 1) ...
Kacise IP66 High Accuracy Wide Pressure Range Digital Pressure Gauge With Data Logger 1. Overview ...
Kacise IP66 High Accuracy Wide Pressure Range Digital Pressure Gauge With Data Logger 1. Overview ...
Kacise IP66 High Accuracy Wide Pressure Range Digital Pressure Gauge With Data Logger 1. Overview ...
Kacise IP66 High Accuracy Wide Pressure Range Digital Pressure Gauge With Data Logger 1. Overview ...