30s±0.1s Lab Glow Wire And Cable Testing Equipment 500-1000°c ±2°c Adjustable
 
Product Description:
Glow wire testing machines are used for the burning test of insulating materials in place for parts of electrical equipment which are not fixed live parts but which provide insulation against electric shock and for the insulation of fixed safety extra-low voltage parts. Imitation products due to fault conditions within the high temperature, such as overload, short circuit and other excessive temperature caused by the fire hazard.
Glow wire test machine is based on IEC335, IEC60695-2-10: 2000 ~ IEC60695-2-13: 2000 "Glow wire / hotline, the basic test methods, glow wire test equipment and general test procedures" and UL 746A, IEC829 DIN695, VDE0471 , GB5169 and other standards require the use of non-flame ignition source program simulation test project.
Glow wire test machine will specify the material (φ4 Ni80 / Cr20) and the shape of the heating wire with a large current heating to the test temperature (500 ºC ~ 1000 ºC) 1min, the required pressure (1.0N) vertical burning test 30s, Ignition and combustion of the solid and other solid combustible materials, ignition temperature (GWIT), flammability and flammability index of the solid insulating material and other solid combustible materials, and the ignitability of the solid insulating material and other solid combustible materials. (GWFI).

 
 
 Technical Parameters:
1, hot wire: φ4mm ± 0.04mm Ni / Cr (80/20) special standard ring, horizontal standing
2, thermocouple: φ0.5mm K-type, imported insulated high temperature armored thermocouple, armored sets of heat 1100 ºC (better than the standard 1050 ºC)
3, Glow wire electric heating temperature: 500 ºC ~ 1000 ºC adjustable
4, temperature fluctuations: <3 ºC
5, test the test pressure: 1.0N ± 0.1N (adjustable)
6, the maximum penetration depth: 7mm ± 0.5mm (adjustable)
7, the test product moving speed: 10mm / s ~ 25mm / s
8, the test time: 30s ± 0.1s (can be preset)
9, ignition paving board thickness: thick 10mm white pine covered wood 12g / m2 ~ 30g / m2
standard silk paper,
200mm ± 5mm from the glowing wire
10, test area volume: ≥ 0.5M3 background black background illumination ≤ 20Lx;
11, Vent hole: φ100mm
12, power supply: 1kVA 220V 50-60Hz

 

Model NoTH-5601 
ControllerTouch screen controller 
Glow wirediameter 4mm±0.04mmNi/Cr(80/20), standard
Temperature range500-1000°C ±2°C continus adjustable
Thermocouplediameter 0.5mm Type K (Ni/Cr - Ni/Al), armored sheath can withstand temperature -40~ 1100°C
Searing Time30s±0.1s (OMRON H5CN-XAN-Z digital timer, 0-9999s, presettable)
Light-off Time0-999.9 s ± 0.1 s, automatic recording, manual timeout
Fireouttime0-999.9 s ± 0.1 s, automatic recording, manual timeout
Ironing into depth7mm ± 0.5 mm
Sample pressure1N ± 0.2 N
Test speed18±3mm/s
Test backgroundblack background
Test processautomatic control, independent convulsions
Chamber≥0.5m
SizeW1000mm×D650mm×H1300mm
ControlUsing single chip microcomputer + touch screen control
Consult standardGB/T5169.10-2006 , IEC60695-2-10:2000,UL746A
Power SupplyAC 220V, 50/60Hz

 
 

 Standard:
 
GB/T5169.10-2006 , IEC60695-11-5:2000,UL746A
 

 





























Hot Products

Please email to Sandy@yuyangtest.com if you need details. BS476 Part 6 Fire Propagation Index Test ...
Please email to Sandy@yuyangtest.com if you need details. ISO 5660 Cone Calorimeter Fire Testing ...
Please email to Sandy@yuyangtest.com if you need details. EN ISO 9239-1 Flooring Radiant Panel Test ...
Please email to Sandy@yuyangtest.com if you need details. EN 13823 SBI Single Burning Items Test ...
Please email to Sandy@yuyangtest.com if you need details. EN ISO1182 Non-Combustibility Test ...
ISO 6722-1 Lab Single Core Wire Fire Testing Equipment with 60V/600V Introductions: Single wire / ...