Fire Retardance Flammability Testing Labs Measuring Heat Release
Properties
International Standards:
Accord with standard:ISO 5660 /ASTM E 1354/GB T16172
ISO 5660 | Reaction to fire tests – Heat release, smoke production and mass
loss rate |
ASTM E 1354 | Standard Test Method for Heat and Visible Smoke Release Rates for
Materials and Products Using an Oxygen Consumption Calorimeter |
BS 476 Pt.15 | Fire tests on building materials and structures – Method for
measuring the rate of heat release of products |
Technology data:
1.) 19inch standard instrument frame,17inch touchable screen pc
insert
2.) cone heater powering rating:5000w,heat output rate:0~100kw/m2
3.) in the range of 50*50mm in the center surface of expose test
sample ,center irradiation error no more than +-2%;
4.) sample size:100mm*100mm*50mm
5.) weight range:0~2000g;precision:0.1g;
6.) ignition system with safe cut device high pressure spark
generator ,automatic positioning
7.) clockwise magnetic oxygen analyzer ,adopt pressure change of
clockwise magnetic to measure the
oxygen density
8.) density range:0-25%,T90<1.5S,zero
drift:<0.5%month,measure data deviation<0.5%month,
9.) non-dispersion infrared co and co2 analyzer:co:0~1%;co2:0~10%;
10.) smoke analyzer use laser system test smoke density,and this
system was made up of photoelectricity two-electrode valve,0.5mv
He-Ne laser,main graphics detector and auxiliary graphics detector
.
11.) exhaust system was made up of fan,hottle,air intake and
exhaust tube of fan and orifice meter.exhaust gas flow 0~50g/s can
be adjusted,precision 0.1g/s.
12.) ring sampler need to be located in the air take tube at
distance of 685mm against hottle ,need to be 12 holes existed in
the sampler in order to homogeneous air flow component.
13.) exhaust gas flow need to be measured by the differencial
pressure of side of hole plate above the fans at the distance of
350mm,the inner diameter of hole plate is 57mm+-1mm;
14.) the temperature of air flow is tested by armored thermocouple
with closed node of diameter 1.6mm,thermocouple need to be located
in 100 mm above the orifice plate;
15.) gas sampling system including ring sampler ,sampling
pump,filter,cold trap,waste water exhaust,water filter and co2
filter;
16.) cold trap:0~5 degree,diaphragm pump,flow rate :261/min,vacuum
degree:700mm Hg,pressure:2.5bar
17.) temperature controller instrument can be automatically adjust
in 0-1000℃,both setted resolution and temperature control precision
is +-2degree,and supply automatic cold end compensator with
thermocouple
18.) use carton foil type heat flow meter ,design range 0~100k
w/m2,diameter of radiation receiving target is 12.5mm,surface with
durable dull black coating.radiation receiving target is cooling
water method. Precision of heat flow meter is 3%,repeatability is
+0-0.5%,attached a adjust report can be date back to NIST.
19.) supply portable cooling water system.when use heat flow meter
,users do not need to tap and equipped with water.
20.) In order to calibrate the response of the whole test system, a
square hole with a square hole and a square yellow copper tube is
used as the calibration burner for measuring the C- coefficient
values.
21.) Data acquisition system should be able to record the oxygen
analyzer, orifice meter, thermocouple and other instruments of the
output.
22.) Equipped with software operating system, the test results
include: heat release rate, gas flow rate, C coefficient, sample
time and time, total oxygen consumption, total smoke, mass loss
rate, total heat release, effective combustion heat, carbon dioxide
emissions and carbon dioxide emissions.
Description:cone calorimeter tester
Soft ware instruction of cone calorimeter:
1.) Set in order to calibrate the sensor, including the oxygen
analyzer, carbon dioxide analyzer, carbon monoxide analyzer, micro
pressure difference sensor, smoke density measurement system,
weighing device, mass flow control single or double point
calibration, in order to obtain the best linear;
2.) C- coefficient calibration, the software can automatically set
the C coefficient measurement of gas flow, such as 1KW, 3KW or 5KW,
the computer system automatically calculates the ISO 5660 C
coefficient and the average C coefficient, and can save the record;
3.) The software can automatically generate the C- coefficient log,
which is convenient for the user to view the history of the cone
calorimeter, and to identify the accuracy and stability of the
system;
4.) The system can automatically calculate the delay time of the
oxygen analyzer, the carbon dioxide analyzer and the carbon
monoxide analyzer, which is convenient for the synchronous
calculation
5.) Status check interface, the working state of the various sensor
components of the instrument can be clear at a glance;
6.) Can record the work value of each sensor, including micro
pressure difference sensor, chimney temperature, oxygen analyzer,
carbon dioxide analyzer, carbon monoxide analyzer;
7.) The report template is EXCELL format, which can display
graphics and numerical model
Technical data:
Sample size | 100*100*50mm |
Weight range | 0-200g |
precision | 0.1g |
Density range | 0-25% |
T90 | <1.5S |
Zero drift | <0.5%month |
Measure data deviation | <0.5%month |