Guangdong Hongtuo Instrument Technology Co,Ltd |
|
Computerized Rubber Moving Die Rheometer (MDR) Testing Machine / Equipment / Instrument / Apparatus / Device / Method
Equipment summary
Rubber Measurement No Rotor Rheometer is widely used in rubber processing industry, rubber quality control and basic research rubber, For optimize formula of rubber provide accurate data, It can accurately measure the scorch time, rheometer time, sulfide index, the maximum and minimum torque and other parameters.
Main functions
Rubber Measurement Rotor Rheometer used monolithic rotor control,
which include: host, temperature measurement, temperature control,
data acquisition and processing, sensors and electrical chains and
other components. These measurements, temperature control circuit
consists of a temperature control device, platinum
resistance,heater composition, capable of automatic tracking power
and ambient temperature changes, automatically correct PID
parameters to achieve fast and accurate temperature control
purposes. Data acquisition system and mechanical linkage to
complete the rubber vulcanization process of force torch signal
automatic detection, automatic real-time display of temperature and
settings. After curing, automatic processing, automatic
calculation, print vulcanization curve and process parameters. Show
curing time, curing power Ju, also has a variety of audible alert.
Rubber Measurement Rotor Rheometer by computer, the computer setting the parameters of the direct control of test parameters rheometer. Display real-vulcanization curve and temperature curve, store test results, different adjustable comparison of test results and in a different color.
Specification
The standard for the implementation of the vulcanization instrument | ASTM D5289-95 ISO6502-1999(E) GB/T16584-1996 |
Temperature range | room temperature to 200 °C |
Temperature measurement accuracy | 100 to 200°C, ≤±0.1°C |
Temperature stability | 100 to 200°C, ≤±0.1°C |
Setting range | 0 to 200°C |
Setting minimum unit | 0.1°C |
Temperature display range | 0 to 200°C |
Display resolution | 0.1°C |
Set the time range | 0 to 200 minutes |
Set the minimum unit | 1 minute |
Automatic range switching | automatic increase according to actual torque |
Power | 50HZ, ~220V±10%, 50±1Hz, must be well grounded |
Compressed air | 0.32Mpa controlled by pneumatic control valve |
Swing frequency | 100 times/minute (about 1.7HZ) |
Ambient temperature | normal temperature -40°C, relative humidity less than 80%, no corrosive gas in the air |
Weight | ≈250kg |
Power | 2KW |
Dimensions | 680×630×1100(mm) |