Chongqing Gold Mechanical&Electrical Equipment Co.,Ltd. |
|
GD-2801C ASTM D5 Grease Penetrometer
Purpose
•GD-2801C Asphalt and lubricating grease Penetrometeris designed and made as per the National Standard GB/T 4509 “Test
Methods for Needle Penetration of Petroleum Asphalt”, GB/T 269
“Test Methods for Cone Penetration of Lubricating Grease and
Petroleum Grease”, and T0604 “Asphalt Penetration Test” in the
Industry Standard JTJ052 Specification and Test Methods of Bitumen
and Bituminous Mixture for Highway Engineering.
•GD-2801C Asphalt and lubricating grease Penetrometeris used to determine penetration of asphalt and lubricating grease.
1.Asphalt: the needle penetration of asphalt is expressed as the depth that
a standard needle of 100±0.05 g vertically penetrates into a sample
without any measurable friction in 5 seconds in the constant
temperature bath of 25°C±0.1°C. It is to evaluate consistency of
asphalt. The larger the penetration is, the less the consistency is
and the softer the asphalt is.
2.Lubricating Grease: the cone penetration of lubricating grease or petroleum grease is
expressed as the depth that a standard cone of 150 g vertically
penetrates into a sample without any measurable friction in 5
seconds in the constant temperature bath of 25°C±0.1°C. It is used
to evaluate the hardness of lubricating grease or petroleum grease.
• The Asphalt and lubricating grease Penetrometer can also be used
for solid granule, powder, colloid and other materials test, as
well as cream, glycine, butter, margarine cream, fermentation, and
other food inspection. It has been widely used in food, highway
construction and other industries.
Technical Parameters
1. Timing device: you can select 5, 8, 10, 12, 30 or 60 s. timing
error is less than ±0.1 s.
2. Depth measurement device: It uses racks to drive raster for
display. It is convenient, rapid, and intuitive, and it can be
started at any position.
3. Penetration accuracy: ±1 penetration;
4. Maximum penetration: 600 penetrations;
5. Heating power: 220 W;
6. Temperature controlling accuracy: 25±0.1°C (the ambient
temperature should not be higher than 20 °C)
7. Constant temperature bath: made of harden glass;
8. Standard needle: 2.5±0.05g; as per GB/T4509, and T0604;
9. Standard cone: 102.5±0.05g; as per GB/T269;
10. Up and down holder: it can be adjusted as your requirements. It
is convenient to let the needle points just contact sample surface.