Standards: IEC60112: 2003, UL746A, ASTM D 3638-92,DIN53480,GB4207

 

The leakage tracking index tester is used to test the electric tracking index (CT1) and Electric tracking resistance index(PT1). When the short circuit current is greater than 0.5A for 2 seconds, the relay will act and cut off the current, indicating that the test product is unqualified;

 

 

Features

 

High quality solenoid valve to accurately control the size of droplet, no need manual adjustment

Drain function to exhaust the air inside the pipeline

Precise leak current meter to judge the current to guarantee accurate CTI

Precisely adjust the sample desk height to adjust the 2 electrode position, make it easier for testing

Standard explosion proof lamp, conform to the safety standard

Sealed exhaust fan guarantee sealing, better exhaust effect

Precise electronic scale to check the action force of electrode to the sample

Quality electronic components guarantee long term working performance

The distance between the electrodes and the height of the tray are adjustable;

Dropping device: no adjustment is required during the test, and the operation is simple

When the short circuit current is greater than 0.5A for 2 seconds, the relay will act and cut off the current, indicating that the test product is unqualified;

 

 

Specifications

 

Reaction force from electrode to the sample1.0±0.05N
Electrode materialcopper + platinum (platinum purity reaches 99% of standard requirements)
Drop time30s ± 0.01s (better than standard by 1 second)
Applied voltage100 ~ 600V (48 ~ 60Hz),the voltage drop does not exceed 10% when the short-circuit current is 1.0 ± 0.0001A (better than the standard 0.1A)
Drop height30 ~ 40mm, 44 ~ 55 drops / 1cm³
Volume of combustion test areaL1050mm × W640mm × H1300mm
backgroundBlack, background illumination ≤20Lux
Equipment dimensionsL1310 * W650 * H1560mm
Exhaust holeØ100mm
Working powerAC220V ± 10%, 10A, 50HZ

 

: :

Hot Products

Specifications Executive standards GBT 1346-2011 Test Method for Water Consumption, Setting Time, ...
It’s a precision spring testing equipment, specially suitable for testing of precise spring, ...
Standards: JB/T 7796,JIS B7738-2001,JJF1134-2005 It’s a precision spring testing equipment, ...
Standards: GB/T1 3927 -92, GB12242-89,ZBJl6006-90,CJ/T3055-95 Model: FMS-30-150A This test bench is ...
Standards GB/T3923.1.2; GB/T39172.3; FZ/T01030; ISO13934.1.2; DIN53838; DIN53861.2; ISO13935.1.2; ...
Features Material: Internal material made of 1.5t white iron steel plate, external SS41 steel plate ...