Allow the test cassette, specimen, buffer, and/or controls to
equilibrate to room temperature (15-30 °C) prior to testing.
1. Collect canine’s fresh whole blood or serum or plasma samples
and ensure that the samples cannot be turbid or precipitated. Only
use clear, non-hemolyzed specimens. Do not leave the specimens at
room temperature for prolonged periods. Serum and plasma specimens
may be stored at 2-8 °C for up to 3 days. For long term storage,
serum and plasma specimens should be kept below -20°C
2. Place the test cassette on a clean and level surface.
3. Hold the dropper vertically and transfer 1 drop of sample
(approximately 10 μl) and immediately add 3 drops of buffer
(approximately 120 μl) into each specimen well (S), then start the
timer. See illustration above.
4. Read the result in 10 minutes. Do not interpret results after 20
minutes.
Note: For Ehrlichia window, if the test line is very strong or even
darker than the control line or the control line disappears, it may
be a strong positive sample other than invalid result. It is
recommended to dilute the sample by 3-5 times with the buffer and
retest with a new test to confirm the result.
INTERPRETATION OF RESULTS
CHW Ag/Ehrlichia Ab Positive: The presence of both C line and T
line
Lyme IgG Positive: Two distinct colored lines appear. One colored
line should be in the control region (C) and another colored line
should be in the IgG region.
Lyme IgM Positive: Two distinct colored lines appear. One colored
line should be in the control region (C) and another colored line
should be in the IgM region.
Lyme IgG and IgM Positive: Three distinct colored lines appear. One
colored line should be in the control region (C) and another two
colored lines should be in the IgG and IgM regions.
Negative: Only clear C line appears.
Invalid: No colored line appears in C zone, no matter whether T
line appears. For Ehrlichia window, if the T line is very strong
but no C line or weak C line, it may be a result of strong positive
sample, it is recommended to dilute the specimen and retest with a
new test to confirm.