Address:
Room 10505, Unit 1, 2th Building GuanTangShengJing , Crossing the Southwest Corner of the South TuanJie Road & DaZhai Road , YanTa District, Xi'an City, Shaan Xi
1. frequency response: 5HZ~1000HZ (3-dB) 2. sensitivity: 20mv/mm/s5% 80HZ, speed of 10mm/s determination in
case 3. magnitude linearity: <3% 4. transverse sensitivity ratio: <5% 5. temperature range: -30~120 ℃ 6. the maximum displacement: 1mm (single peak) 7. direction: installation requires below to the sensor is 0 ° Vertical type02.5vertical Horizontal type902.5horizontal 8. Output impedance900. 9. insulation resistance: >2M. 10. The output polarity: floating output. When the sensor base by
an impact force, 1# core output (+), 2# (-) for the core output. Access oscilloscope, 1# core connected
core pole, 2# core grounding, oscilloscope display the first peak.