Product code | 221142 | 221143 | 221162 | 221163 |
Model | ICL300A | ICL300B | ICL310A | ICL310B |
Type | Desktop type | Installation type | Desktop type | Installation type |
System | Internal chamber wall heating natural convection system |
Performance*1 |
Operating temp. range | Room temp.5 to 70 deg.C |
Temp. adjustment accuracy | /-0.5 deg.C (at 37 deg.C) |
Temp. distribution accuracy | /-1.0 deg.C (at 37 deg.C) |
Configuration |
Internal material | Stainless steel plate |
External material | Chrome free galvanized steel plate |
Heat insulator | Glass wool |
Door | Front opening system |
Shelf | 2 as standard, Deep punching shelf board with a pull-handle
(height: 25 mm) |
Number of shelves | Five at 50 mm pitch |
Packing | Magnet packing |
Heater | 120V 145W | 240V 145W |
Leg | Rubber leg | Caster wheels | Rubber leg | Caster wheels |
Controller |
Temp. control system | PID control with a micro computer |
Temp. setting system | Digital setting display |
Operation function | Fixed operation & timer operation (quick auto stop, auto stop,
auto start) |
Control sensor | K-thermocouple |
Function |
Safety functions | Over current electric leakage breaker, automatic overheat
prevention function, bimetal overheat preventive device (manual
reset) |
Standards |
Chamber internal dimensions and capacity | 350 x 300 x 330mm (W x D x H), 34.7L |
External dimensions*2 | 450 x 400 x 600mm (W x D x H) |
Power source | Multi power source AC100 to 120V, 1.2A | Multi power source, single phase AC200 to 240V, 0.6A |
|